Calibration
4-49
Camera of Brivo OEC 865
5075982 CAMERA 865, GigE Camera GE1020B
Camera ASM 6491850, GigE Camera GE1020C
The secure screw of 5085003 is 1/16 inch HEX.
Don’t touch the iris internal secure screw which can lead to iris adjustment ring disabled.
6.
Make an exposure to make sure the entrance dose rate:
Brivo OEC 715/Brivo OEC 785(I.I PN: 5075001):
4.1±0.3mR/min (35.92±2.63uGy/min or
2155±157uGy/hr)
Brivo OEC 715/785/865 (I.I PN: 5085917)
:
3.1±0.3mR/min (27.16±2.63uGy/min or
1629±157uGy/hr)
Brivo OEC 865(I.I PN: 5075008):
3.1±0.3mR/min (27.16±2.63uGy/min or 1629±157uGy/hr)
Camera Iris
Adjustment Ring
Iris internal
secure Screw
on the back
Camera Iris
Adjustment Ring
Camera Iris
Secure Screw