Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
3-52
CAUTION
CAUTION
CAUTION
When the laser aimer is attached to the C-Arm, releasing the orbital or rotational brake
could cause the C-Arm to move. Make sure to hold the C-Arm to control its movement
whenever a brake is released.
1.
Place the laser aimer’s attachment handle in the open position.
2.
Move the laser aimer flush with the Image Intensifier, but slightly off center.
3.
Center the laser aimer with the relation to the Image Intensifier. As you center the laser aimer, the
attachment hooks catch on the edge of the image intensifier.
4.
Turn the attachment handle inward until the handle is completely
closed and the laser aimer is securely attached to the image
intensifier. Insert the metal bolt to the hole of the handle to lock.
Do not grasp the Laser Aimer's handle to position the C-Arm or to move the system.
Pulling the handle may cause the Laser Aimer to release from the image intensifier and
fall.
5.
Use the image intensifier handle to position the C-Arm or to move the system to the desired location.
6.
After attaching the laser aimer, place the C-Arm in the position it will be used during the procedure.
7.
Turn the laser aimer ON and adjust the beam alignment
knobs until the laser beam is centered on the intersection of
the cross-hair on the Monoblock cover.
If the laser is ON, do not look directly into the laser aimer’s aperture.
8.
Functional check: Using fluoroscopy, verify that the cross-hair in the fluoro image, at the center of the
area of interest.
The location where the laser beam marks the patient corresponds to the center of the
area of interest, indicated by the cross-hair in the fluoro image.
Perform procedure below to remove the laser cover from the laser aimer:
1.
Switch OFF the power to the laser aimer.
2.
Grasp the two knobs on the laser cover.
3.
Press down and rotate the cover (approximately 10 degrees counterclockwise) until the knobs align
with the notches on the housing.
4.
Lift off the cover and resume laser aimer operation.
Summary of Contents for Brivo OEC 715
Page 2: ......
Page 19: ...Chapter1 Introduction and Safety...
Page 41: ...Introduction and Safety 23 46 54 20 18 26...
Page 55: ...Chapter2 System Overview...
Page 137: ...Chapter3 Installation...
Page 212: ...Chapter4 Calibration...
Page 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Page 284: ...Chapter5 Software...
Page 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Page 335: ...Chapter6 Troubleshooting...
Page 408: ...Chapter7 Replacement...
Page 418: ...Replacement 7 11 166...
Page 488: ...Chapter8 Periodic Maintenance...
Page 502: ...Periodic Maintenance 8 15...
Page 505: ...Chapter9 Technical Reference...
Page 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Page 526: ...11 Appendix System Schematics...