Calibration
4-57
2.
Login Administration interface by service authority, select Calibration Tab, click “shading
Calibration” button.
3.
In shading calibration interface, click each Mag calibration buttons, shading correction
calibration Wizard will be prompted. According to the Wizard complete the calibration.
4. Redo the Shading Correction Calibration Check to ensure the system is in good brightness