Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
3-18
3.5.
System Setup
3.5.1.
Power Supply Adaption
If your power supply is 100V/110V/115V/120V/200V/220V/230V/240V, please follow steps below to
change power supply. Otherwise, go to section3.5.2.
1.
Make sure that workstation rear cover was removed.
2.
Check the breaker, power cable and plug are proper and relative to the power supply voltage.
3.
Input voltage option plug. The short terminator is used for setting input voltage. Upper option plug is
for 110V (100V/110V/115V/120V), under the 110V plug is the 220V plug (200V/220V /230V/240V):
4.
Change the bridge connectors of the transformer if need.
AC 100/110/115/120
AC 200/220/230/240V
110V Jumper
220V Jumper