Periodic Maintenance
8-5
mA Accuracy (
±
10%)
Low Dose:
Normal Fluoro:
0.1mA@80kVp: , 4.0mA@110kVp: ,
HLF:
4.0mA(Actual 4.5)@110KVp: , (Dose Group: Normal 50 and HLF 50)
4.0mA(Actual 6.0)@110kVp: , (Dose Group: Normal 50 and HLF 100)/(Dose Group: Normal 88
and HLF 100)
Digital Spot:
3.2mA(Actual 6.4)@80kVp: ,
4.0mA(Actual 10)@110kVp: , ( For 100V/110V/120V input)
4.0mA(Actual 16)@110KVp: , (For 200V/220V/230V/240V input)
Radiography:
For power input: 100V/110V/120V:
40KVp@40mAs (Actual 10mA): , 110KVp@40mAs (Actual 10mA): ,,
For power input: 200V/220V/230V/240V
40KVp@80mAs (Actual 20mA): , 110KVp@40mAs (Actual 20mA): ,
mAs Accuracy(
±
10%)
40kV, 40mAs: , 110kV, 1mAs: , 110kV, 2mAs: , 110KV, 40mAs: ,
Image Quality Check
Image Intensifier Field Size Check
Mag Mode
NORM
(
210
±
10
)
MAG1
(
155
±
8
)
MAG2
(
115
±
8
)
Displayed Image Diameter (mm)
Auto Technique Tracking and Image Resolution Check
Auto Technique Tracking
Image Resolution
Copper
Filters
(mm)
Specification
(kVp)
Measured
(kVp)
Adjustment
done?
Mag Mode
(Field Size)
Specification
(LP/mm)
Measured
(LP/mm)
Adjustment
done?
715/78
5
(II PN:
50750
01)
715/785/865
(I.I PN: 5085917)
865
(I.I
PN:
5075
008)
718/785
865
1
64
±
4
62
±
4
62
±
4
Yes No
NORM (9”)
≥
2.0
≥
2.2
Yes No
2
74
±
4
72
±
4
72
±
4
Yes No
MAG1 (6”)
≥
2.6
≥
3.0
Yes No
3
82
±
4
80
±
4
80
±
4
Yes No
MAG2 (4.5”)
≥
3.1
≥
3.5
Yes No
Shading Correction Check
Pass
Fail
Dose Check
Dose rate Check
Measured as found:
mR/min
No adjustment necessary:
Measured if
adjusted:
mR/min
715/785(II PN: 5075001)
715/785/865 (I.I PN: 5085917)
865(I.I PN: 5075008)
4.1
±
0.3mR/min
3.1±0.3mR/min
3.1±0.3mR/min
DAP Accuracy
Check
Pass Fail (Need Calibration)
DAP Calibration Done
Reporting
Install Equipment Covers
Yes
No
Reporting complete
Yes
No
Periodic maintenance of the system is required to maintain the system lifecycle. The first periodic
maintenance shall be implemented after first 13 months’ use, then it is suggested to do once per year by
GE qualified service engineer.
Summary of Contents for Brivo OEC 715
Page 2: ......
Page 19: ...Chapter1 Introduction and Safety...
Page 41: ...Introduction and Safety 23 46 54 20 18 26...
Page 55: ...Chapter2 System Overview...
Page 137: ...Chapter3 Installation...
Page 212: ...Chapter4 Calibration...
Page 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Page 284: ...Chapter5 Software...
Page 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Page 335: ...Chapter6 Troubleshooting...
Page 408: ...Chapter7 Replacement...
Page 418: ...Replacement 7 11 166...
Page 488: ...Chapter8 Periodic Maintenance...
Page 502: ...Periodic Maintenance 8 15...
Page 505: ...Chapter9 Technical Reference...
Page 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Page 526: ...11 Appendix System Schematics...