System Overview
2-17
Mainframe Control Board
The MCB (Mainframe Control Board) is located in C-Arm mainframe and is the main control unit in mainframe. It is
responsible for five functions.
-
X-ray control: KV, mA and exposure control interfacing with collimator board, filament board, and KV control
board
-
Image chain control: control camera and image intensifier
-
User interface: key input and LED display interfacing with display board
-
System monitor: check various error and warning
-
Communication with workstation
The structure of MCB interface is illustrated in the following diagram.
Functional Block
MCB consists of 10 function blocks, which are illuminated as follows: