Installation
3-29
Add and edit screen of Storage & Commitment Servers
Storage & Commitment Servers
Function Check:
Press Image Directory
button.
Select DCIOM Storage Tab.
Click “Export” Tab.
In Store to pull down menu,
select available Select the
DICOM storage device.
Select one or more preview
images you want to archive.
Press and hold CTRL key on the
keyboard to select multiple
images from the previews in the
directory. Click “send” button.
Verify that the images are
copied to the storage server
successfully after the copy
progress complete.
All images successfully sent to
DICOM storage server will be
marked as “S”, if the “Enable
storage commitment” check