Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
6-22
Exception
ID
Error
Warning
Exception
Description
(In system Log)
Message
(Displayed in User Interface)
Diagnosis Reference
for service.
0202501
Error
Allocate
memory failed
Image acquisition exception.
System can’t take exposure.
The mainframe stopped.
Please restart the system. If
this issue persists, please call
for service.
End capture failed.
1. If Gige camera is not
working correctly, please
change camera.
2. If LED on ethnet interface
is not twinkle, please insert
the network cable into
ethnet interface again or
replace another network
cable.
3. If all above methods can’t
fix the exception, please re-
install software.
0202502
Warning
Read shading
template failed
Shading parameter loading
failure. Image quality may be
impacted. Please restart the
system. If this issue persists,
please call for service.
Template file are not
deployed corrected.
0202503
Warning
Write shading
template failed
Shading parameter saves
exception. The parameter was
not saved. Please re-calibrate
shading.
Disk is full or directory has
no permission to write.
0202504
Warning
Load shading
calibration
failure
Load shading calibration data
failure. Image quality may be
affected. Please restart the
system. If this issue persists,
please call for service.
0203501
Error
Allocate
memory failed
Run out of memory. Please
restart the system. If this issue
persists, please call for service.
Check memory using state
0204501
Error
Open XDisplay
failed
Failed to find the display
device. Images can’t be
displayed. Please restart the
system. If this issue persists,
please call for service.
make sure monitor
connection, if display driver
installed correctly
0204502
Error
Create display
context failed
Display driver exception. Image
can’t be displayed. Please
restart the system. If this issue
persists, please call for service.
make sure display driver
installed correctly
0204503
Error
Create display
window failed
Display driver exception. Image
can’t be displayed. Please
make sure display driver
installed correctly
Summary of Contents for Brivo OEC 715
Page 2: ......
Page 19: ...Chapter1 Introduction and Safety...
Page 41: ...Introduction and Safety 23 46 54 20 18 26...
Page 55: ...Chapter2 System Overview...
Page 137: ...Chapter3 Installation...
Page 212: ...Chapter4 Calibration...
Page 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Page 284: ...Chapter5 Software...
Page 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Page 335: ...Chapter6 Troubleshooting...
Page 408: ...Chapter7 Replacement...
Page 418: ...Replacement 7 11 166...
Page 488: ...Chapter8 Periodic Maintenance...
Page 502: ...Periodic Maintenance 8 15...
Page 505: ...Chapter9 Technical Reference...
Page 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Page 526: ...11 Appendix System Schematics...