Periodic Maintenance
8-17
8.8.4.
Shading Correction Check
Refer to 4.6.9 to perform shading correction check. Perform calibration if needed.
8.9.
Dose Check
8.9.1.
Dose Rate Check
1.
Using a dosimeter for the measurements. Place the dosimeter probe on the I.I. grid center. Use 2mm
copper to cover the beam path on the X-ray tube. Take exposure in Auto fluoro mode. Dose rate
should be within:
Brivo OEC 715/785 (I.I PN: 5075001):
4.1±0.3mR/min (35.92±2.63uGy/min or 2155±157uGy/hr)
Brivo OEC 715/785/865 (I.I PN: 5085917)
:
3.1±0.3mR/min (27.16±2.63uGy/min or 1629±157uGy/hr)
Brivo OEC 865 (I.I PN: 5075008):
3.1±0.3mR/min (27.16±2.63uGy/min or 1629±157uGy/hr)
2.
Indicate if adjustments were made on the Checklist.
8.9.2.
DAP Accuracy Check
Refer to 4.7 to perform DAP accuracy check. Perform the DAP calibration, if needed.
The field size in the form should be measured by field size measure tool.
8.10.
Reporting
8.10.1.
Install Equipment Covers
Install all equipment covers you removed earlier. Make sure all air filters are clean and properly attached
to the covers.
8.10.2.
Complete PM Comments
Write any comments important to the PM procedure in the Comments section of the PM Checklist.
Summary of Contents for Brivo OEC 715
Page 2: ......
Page 19: ...Chapter1 Introduction and Safety...
Page 41: ...Introduction and Safety 23 46 54 20 18 26...
Page 55: ...Chapter2 System Overview...
Page 137: ...Chapter3 Installation...
Page 212: ...Chapter4 Calibration...
Page 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Page 284: ...Chapter5 Software...
Page 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Page 335: ...Chapter6 Troubleshooting...
Page 408: ...Chapter7 Replacement...
Page 418: ...Replacement 7 11 166...
Page 488: ...Chapter8 Periodic Maintenance...
Page 502: ...Periodic Maintenance 8 15...
Page 505: ...Chapter9 Technical Reference...
Page 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Page 526: ...11 Appendix System Schematics...