Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
7-58
Main Power on Breaker
Brivo OEC715/785/865 provides 110V breaker (P/N: 5075974) and 220V breaker (P/N: 5331605).
1.
Remove the rear cover of the Workstation. Unplug the cable connected to the circuit breaker.
2.
Loose the screws and remove the breaker. Then replace a new breaker.
3.
Connect the cables to the breaker and reassemble the parts.
4.
Power on the system. Verify that the Workstation and C-Arm complete software boot and
initialization with no errors reported on their displays.
C-Arm Power Filter (P/N: 5331600)
There are 2 power filters in Brivo OEC715/785/865. One is behind the mainframe package box, the other
is in Workstation.
Power filter in Mainframe
1.
Remove the rear cover of the C-Arm console
first.
2.
Open the mainframe package box to access the
power filter.
3.
Unplug the cable connected to the filter.
4.
Loose the screws and remove the filter. Replace
a new part and connect the cable according to
the schematic on the filter.
5.
Reassemble the new parts. Power on the
system.
6.
Verify that the Workstation and C-Arm complete
software boot and initialization with no errors
reported on their displays.
Summary of Contents for Brivo OEC 715
Page 2: ......
Page 19: ...Chapter1 Introduction and Safety...
Page 41: ...Introduction and Safety 23 46 54 20 18 26...
Page 55: ...Chapter2 System Overview...
Page 137: ...Chapter3 Installation...
Page 212: ...Chapter4 Calibration...
Page 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Page 284: ...Chapter5 Software...
Page 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Page 335: ...Chapter6 Troubleshooting...
Page 408: ...Chapter7 Replacement...
Page 418: ...Replacement 7 11 166...
Page 488: ...Chapter8 Periodic Maintenance...
Page 502: ...Periodic Maintenance 8 15...
Page 505: ...Chapter9 Technical Reference...
Page 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Page 526: ...11 Appendix System Schematics...