Guidance and manufacturer’s declaration - Electromagnetic Immunity
The
Brivo OEC 715/785/865
Mobile C-Arm X-Ray Product is intended to use in the specified
electromagnetic environment. The purchaser or user of the
Brivo OEC 715/785/865
Mobile C-Arm X-Ray
Product should assure that it is used in an electromagnetic environment as described below:
IEC 60601
Test Level
Compliance Level
Electromagnetic Environment
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms , 6Vrms
at ISM band
150GHz to
80MHz
3V /m 80MHz to
2.5GHz (Ed3.0)
3V/m 80MHz to
2.7GHz, 28V/m
(Ed4.0)
3Vrms , 6Vrms at
ISM band 150GHz
to 80MHz
3V/m 80MHz to
2.5GHz (Ed3.0)
3V/m 80MHz to
2.7GHz, 28V/m
(Ed4.0)
Portable and mobile RF communications
equipment are used no closer to any part of
the
Brivo OEC 715/785/865
Mobile C-Arm X-
Ray Product, including cables, than the
recommended separation distance calculated
from the equation appropriate for the
frequency of the transmitter.
Recommended separation distance
P
=
d
]
3
5
,
3
[
150 kHz to 80 MHz
P
=
d
]
3
5
,
3
[
80 MHz to 800 MHz
P
=
d
]
3
7
[
800 MHz to 2.5 GHz
The formulas are applicable for Ed3.0 only.
Note: P is the power rating of the transmitter
in watts (W) according to the transmitter
manufacturer and d is the recommended
separation distance in meters (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey
a
, are less than the compliance level in each
frequency range
b
.
Interference may occur in the vicinity of
equipment marked with the following symbol:
a: Field strengths from fixed transmitters, such as base stations for cellular telephones and land mobile radios, amateur
radio, AM and FM radio broadcast, and TV broadcast cannot be estimated accurately. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be performed. If the measured field
strength exceeds the RF compliance level above, observe the Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product to
verify normal operation in each use location. If abnormal performance is observed, additional measures may be
necessary, such as re-orienting or relocating the [EQUIPMENT and/or SYSTEM].
b: Over the frequency range 150 kHz to 80 MHz, field strengths are less than 3 V/m.
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
Note3: These are guidelines. Actual conditions may vary.
Summary of Contents for Brivo OEC 715
Page 2: ......
Page 19: ...Chapter1 Introduction and Safety...
Page 41: ...Introduction and Safety 23 46 54 20 18 26...
Page 55: ...Chapter2 System Overview...
Page 137: ...Chapter3 Installation...
Page 212: ...Chapter4 Calibration...
Page 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Page 284: ...Chapter5 Software...
Page 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Page 335: ...Chapter6 Troubleshooting...
Page 408: ...Chapter7 Replacement...
Page 418: ...Replacement 7 11 166...
Page 488: ...Chapter8 Periodic Maintenance...
Page 502: ...Periodic Maintenance 8 15...
Page 505: ...Chapter9 Technical Reference...
Page 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Page 526: ...11 Appendix System Schematics...