Troubleshooting
6-33
Exception
ID
Error
Warning
Exception
Description
(In system Log)
Message
(Displayed in User Interface)
Diagnosis Reference
0307510
Error
Measurement
module
unregister
servant
exception
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service.
1. Restart the system.
2. If not work, re-install
Workstation software.
0307511
Warning
Init
measurement
controller failed
Measurement init exception.
Please restart the system. If
this issue persists, please call
for service.
1. Restart the system.
2. If not work, re-install
Workstation software.
0307512
Warning
Parse image or
overlay
exception
Parse image or overlay
exception. Can’t display the
image or overlay. Please
restart the system. If this issue
persists, please call for service.
1. Restart the system.
2. If not work, re-install
Workstation software.
0307513
Warning
Stop bundle
exception
Close Measurement exception.
Please restart the system. If
this issue persists, please call
for service.
1. Restart the system.
2. If not work, re-install
Workstation software.
0307514
Warning
Start
measurement
exception
Open Measurement exception.
Please restart the system. If
this issue persists, please call
for service.
0307515
Warning
Stop
measurement
exception
Close Measurement exception.
Please restart the system. If
this issue persists, please call
for service.
0307516
Warning
Get string failed Loading message fail. Please
restart the system. If this issue
persists, please call for service.
0307517
Warning
Get resource
bundle failed
Loading message fail. Please
restart the system. If this issue
persists, please call for service.
0307518
Error
Measurement
module get
OSGI service
exception
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service.
0307519
Warning
Stop
application
exception
Close Measurement exception.
Please restart the system. If
this issue persists, please call
for service.
Summary of Contents for Brivo OEC 715
Page 2: ......
Page 19: ...Chapter1 Introduction and Safety...
Page 41: ...Introduction and Safety 23 46 54 20 18 26...
Page 55: ...Chapter2 System Overview...
Page 137: ...Chapter3 Installation...
Page 212: ...Chapter4 Calibration...
Page 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Page 284: ...Chapter5 Software...
Page 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Page 335: ...Chapter6 Troubleshooting...
Page 408: ...Chapter7 Replacement...
Page 418: ...Replacement 7 11 166...
Page 488: ...Chapter8 Periodic Maintenance...
Page 502: ...Periodic Maintenance 8 15...
Page 505: ...Chapter9 Technical Reference...
Page 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Page 526: ...11 Appendix System Schematics...