Brivo OEC 715/785/865/865 Mobile C-Arm X-Ray Product Service Manual
2-62
connected means FOOT BUTTON 1 has
been pressed
FOOT_BUTTON2
Button 2
Passive
Open/Close J1-A19 FOOT_BUTTON2 & FOOT_COM is
connected means FOOT BUTTON 2 has
been pressed
FOOT_COM
Common
termination
Passive
-
J1-B32 -
Emergency switch, located on the control panel of C-arm mainframe, stops all motorized motion immediately,
interrupts any X-rays in progress, and prevents further X-rays until you reset the system. At any time, if the
emergency switch is pushed down, the X-ray exposure shall be stopped immediately. System shall trigger the
related error message.
Name
Description
Attribution
Scale
Comment
EMERGENCY_SWT
One termination of
Emergency
switch1
Passive
Open/Clos
e
EMERGENCY_SWT & REF_EMERGENCY_SWT is
connected means EMERGENCY_SWT has pressed
REF_
EMERGENCY_SWT
The other
termination of
Emergency
switch1
Passive
-
-
EMERGENCY_SWT2
One termination of
Emergency
switch2
Passive
Open/Clos
e
EMERGENCY_SWT2 & REF_EMERGENCY_SWT2 is
connected means EMERGENCY_SWT2 has pressed
REF_
EMERGENCY_SWT2
The other
termination of
Emergency
switch2
Passive
-
-
MF_AC28V_L
-
-
-
-
MF_AC28V_N
-
-
-
-
Summary of Contents for Brivo OEC 715
Page 2: ......
Page 19: ...Chapter1 Introduction and Safety...
Page 41: ...Introduction and Safety 23 46 54 20 18 26...
Page 55: ...Chapter2 System Overview...
Page 137: ...Chapter3 Installation...
Page 212: ...Chapter4 Calibration...
Page 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Page 284: ...Chapter5 Software...
Page 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Page 335: ...Chapter6 Troubleshooting...
Page 408: ...Chapter7 Replacement...
Page 418: ...Replacement 7 11 166...
Page 488: ...Chapter8 Periodic Maintenance...
Page 502: ...Periodic Maintenance 8 15...
Page 505: ...Chapter9 Technical Reference...
Page 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Page 526: ...11 Appendix System Schematics...