Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
3-70
3.7.11.
Mono LCD monitor installation
Refer to Chapter 7 to install the Mono monitor.
3.7.12.
Software Option Installation
Refer to Chapter 5 to install software option function. It can be implemented in 3.5.5 System
Configuration together with product model configuration.
3.7.13.
Room Interface (RIF)
Room Interface (RIF) is an interface interactive with operation room or some peripheral equipment. The
lamps in operation rooms such as Room In Use indicator or X-Ray On indicator can be connected to
system to warn user, injector can be controlled by room interface if necessary, also the exposure can be
restrained of door status in case of excessive radiation to user, all functions in RIF are not mandatory
when operating.
System provides a HDB15 connector as room interface output/Input to connect each indicators, on the
back of workstation. And door switch work mechanism software setup interface in workstation software.
Summary of Contents for Brivo OEC 715
Page 2: ......
Page 19: ...Chapter1 Introduction and Safety...
Page 41: ...Introduction and Safety 23 46 54 20 18 26...
Page 55: ...Chapter2 System Overview...
Page 137: ...Chapter3 Installation...
Page 212: ...Chapter4 Calibration...
Page 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Page 284: ...Chapter5 Software...
Page 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Page 335: ...Chapter6 Troubleshooting...
Page 408: ...Chapter7 Replacement...
Page 418: ...Replacement 7 11 166...
Page 488: ...Chapter8 Periodic Maintenance...
Page 502: ...Periodic Maintenance 8 15...
Page 505: ...Chapter9 Technical Reference...
Page 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Page 526: ...11 Appendix System Schematics...