Troubleshooting
6-7
Exception
ID
Error
Warning
Exception
Description
(In system Log)
Message
(Displayed in User Interface)
Diagnosis Reference
0103006
Warning
Leaves angle
feedback SPI
channel is
abnormal
Collimator leaves rotation
monitoring exception. Leaves
can’t rotate. Please restart the
system. If this issue persists,
please call for service.
MCU angle width A/D
failure, MCB may be
abnormal
0103007
Warning
Iris channel SPI
bus Tx is time
out
Collimator iris control
exception. Iris can’t open or
close. Please restart the
system. If this issue persists,
please call for service.
1. MCB may be abnormal
2. Mother Board may be
abnormal
3. CCB may be abnormal
0103008
Warning
Iris channel SPI
bus Rx is time
out
Collimator iris control
exception. Iris can’t open or
close. Please restart the
system. If this issue persists,
please call for service.
1. MCB may be abnormal
2. Mother Board may be
abnormal
3. CCB may be abnormal
0103009
Warning
Leaves
open/close
channel SPI bus
Tx is time out
Collimator leaves open/close
control exception. Leaves can’t
open or close. Please restart
the system. If this issue
persists, please call for service.
Leaves open/close channel
SPI (MCU) bus Tx is time out
1. MCB may be abnormal
2. Mother Board may be
abnormal
3. CCB may be abnormal
0103010
Warning
Leaves
open/close
channel SPI bus
Rx is time out
Collimator leaves open/close
control exception. Leaves can’t
open or close. Please restart
the system. If this issue
persists, please call for service.
Leaves open/close channel
SPI (MCU) bus Rx is time out
1. MCB may be abnormal
2. Mother Board may be
abnormal
3. CCB may be abnormal
0103011
Warning
Leaves rotation
channel SPI bus
Tx is time out
Collimator leaves rotation
control exception. Leaves can’t
rotate. Please restart the
system. If this issue persists,
please call for service.
Leaves rotation channel SPI
(MCU) bus Tx is time out
1. MCB may be abnormal
2. Mother Board may be
abnormal
3. CCB may be abnormal
0103012
Warning
Leaves rotation
channel SPI bus
Rx is time out
Collimator leaves rotation
control exception. Leaves can’t
rotate. Please restart the
system. If this issue persists,
please call for service.
Leaves rotation channel SPI
(MCU) bus Rx is time out
1. MCB may be abnormal
2. Mother Board may be
abnormal
3. CCB may be abnormal
0103800
Warning
Collimator is
not calibrated.
Collimator not calibrated. Iris
and leaves can’t be
manipulated. Please call
service to calibrate the
collimator.
1. Calibrate collimator
Summary of Contents for Brivo OEC 715
Page 2: ......
Page 19: ...Chapter1 Introduction and Safety...
Page 41: ...Introduction and Safety 23 46 54 20 18 26...
Page 55: ...Chapter2 System Overview...
Page 137: ...Chapter3 Installation...
Page 212: ...Chapter4 Calibration...
Page 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Page 284: ...Chapter5 Software...
Page 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Page 335: ...Chapter6 Troubleshooting...
Page 408: ...Chapter7 Replacement...
Page 418: ...Replacement 7 11 166...
Page 488: ...Chapter8 Periodic Maintenance...
Page 502: ...Periodic Maintenance 8 15...
Page 505: ...Chapter9 Technical Reference...
Page 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Page 526: ...11 Appendix System Schematics...