Troubleshooting
6-15
Exception
ID
Error
Warning
Exception
Description
(In system Log)
Message
(Displayed in User Interface)
Diagnosis Reference
0117005
Warning
KV feedback
channel SPI is
abnormal
Execution exception. Please
restart the system. If this issue
persists, please call for service.
1.MCB may be abnormal
0117006
Warning
MA feedback
channel SPI is
abnormal
Execution exception. Please
restart the system. If this issue
persists, please call for service.
1.MCB may be abnormal
0117007
Warning
Little MA
feedback
channel SPI is
abnormal
Execution exception. Please
restart the system. If this issue
persists, please call for service.
1.MCB may be abnormal
0117008
Warning
KV feedback
channel SPI Tx
time out
Execution exception. Please
restart the system. If this issue
persists, please call for service.
1.MCB may be abnormal
0117009
Warning
KV feedback
channel SPI Rx
time out
Execution exception. Please
restart the system. If this issue
persists, please call for service.
1.MCB may be abnormal
0117010
Warning
Preheat MA
feedback
channel SPI Tx
time out
Execution exception. Please
restart the system. If this issue
persists, please call for service.
1.MCB may be abnormal
0117011
Warning
Preheat MA
feedback
channel SPI Rx
time out
Execution exception. Please
restart the system. If this issue
persists, please call for service.
1.MCB may be abnormal
0117012
Warning
Little MA
feedback
channel SPI Tx
time out
Execution exception. Please
restart the system. If this issue
persists, please call for service.
1.MCB may be abnormal
0117013
Warning
Little MA
feedback
channel SPI Rx
time out
Execution exception. Please
restart the system. If this issue
persists, please call for service.
1.MCB may be abnormal
0117014
Warning
MAS output is
over limit
Execution exception. Please
restart the system. If this issue
persists, please call for service.
1. MCB maybe abnormal
2.mA feedback is abnormal,
recheck mA accuracy.
0117800
Warning
Preheat MA is
not calibrated
System execution exception.
Please try again. If this issue
persists, please call for service.
N/A
0118001
Warning
Xray leakage is
detected when
there is no
exposure
Execution exception. Please
restart the system. If this issue
persists, please call for service.
Maybe Monoblock
grounding issue.
Summary of Contents for Brivo OEC 715
Page 2: ......
Page 19: ...Chapter1 Introduction and Safety...
Page 41: ...Introduction and Safety 23 46 54 20 18 26...
Page 55: ...Chapter2 System Overview...
Page 137: ...Chapter3 Installation...
Page 212: ...Chapter4 Calibration...
Page 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Page 284: ...Chapter5 Software...
Page 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Page 335: ...Chapter6 Troubleshooting...
Page 408: ...Chapter7 Replacement...
Page 418: ...Replacement 7 11 166...
Page 488: ...Chapter8 Periodic Maintenance...
Page 502: ...Periodic Maintenance 8 15...
Page 505: ...Chapter9 Technical Reference...
Page 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Page 526: ...11 Appendix System Schematics...