Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
6-16
Exception
ID
Error
Warning
Exception
Description
(In system Log)
Message
(Displayed in User Interface)
Diagnosis Reference
0118800
Warning
Exposure is
disabled
because the
room door is
opened
Room door open. Exposure is
disabled. Please close the door
to enable exposure.
N/A
0118801
Warning
Exposure is
automatically
terminated as
the limit of
9'30" for one
shot is reached
Limit of maximum 9'30" for one
shot reached. Exposure is
terminated.
N/A
0118802
Warning
Accumulated
exposure
display time is
longer than
99.9 minutes.
Mainframe
can’t display
rightly
accumulated
exposure time.
Accumulated fluoro time
longer than 99.9 minutes. This
exceeds mainframe fluoro time
display capacity. Please press
and hold Alarm key to reset the
accumulated exposure time.
N/A
0118803
Warning
9'30" exposure
time limit will
be reached in
30"
Limit of maximum 9'30" for one
shot will be reached in 30s.
Exposure will be terminated
then.
N/A
0119500
Warning
Priority II
message buffer
overflow
Mainframe buffer overflow. The
mainframe stopped. Please
restart the system. If this issue
persists, please call for service.
MCU abnormal, MCB may
be abnormal.
STOP0032
768
Warning
WKS error
before MF boot
up
Just display on Mainframe
Control Panel
There is calibration data in
MCB, but WKS didn't
configure product model
Imaging
0205501
Error
Allocate
memory failed
Run out of memory. Please
restart the system. If this issue
persists, please call for service.
Check memory using state
0205502
Error
Allocate shared
memory
segment failed
Memory exception. Please
restart the system. If this issue
persists, please call for service.
Check if there exists some
unwanted share memory in
the system, or check if
software is installed
correctly.
Summary of Contents for Brivo OEC 715
Page 2: ......
Page 19: ...Chapter1 Introduction and Safety...
Page 41: ...Introduction and Safety 23 46 54 20 18 26...
Page 55: ...Chapter2 System Overview...
Page 137: ...Chapter3 Installation...
Page 212: ...Chapter4 Calibration...
Page 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Page 284: ...Chapter5 Software...
Page 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Page 335: ...Chapter6 Troubleshooting...
Page 408: ...Chapter7 Replacement...
Page 418: ...Replacement 7 11 166...
Page 488: ...Chapter8 Periodic Maintenance...
Page 502: ...Periodic Maintenance 8 15...
Page 505: ...Chapter9 Technical Reference...
Page 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Page 526: ...11 Appendix System Schematics...