Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
6-66
Exception
ID
Error
Warning
Exception
Description
(In system Log)
Message
(Displayed in User Interface)
Diagnosis Reference
4. PDU may be abnormal.
0408007
Warning
Secondary
voltage
underload
Secondary voltage underload.
It may result in exposure
exception. Please restart the
system. If this issue persists,
please call for service.
1. Main transformer
primary input power supply
may be abnormal.
2. 28V output of main
transformer may be
abnormal.
3. Adjust R52 follow SM
chapter 4.9
4. PDU may be abnormal.
0408008
Warning
PDU not detect
IPC 12V
Power monitoring exception. If
this issue persists, please call
for service.
1. 12V output of WKS may
be abnormal.
2. 12 V input for PDU may
be abnormal.
0408009
Error
Open serial port
failed
Workstation and peripheral
interface communicaiton
exception. Please restart the
system. If this issue persists,
please call for service.
Serial port of WKS abnormal
1. WKS serial port for UPS
may be abnormal.
2.Reinstall WKS software
0408010
Warning
Write serial port
failed
Workstation and peripheral
interface communicaiton
exception. Please restart the
system. If this issue persists,
please call for service.
Serial port of WKS abnormal
1. WKS serial port for UPS
may be abnormal.
2.Reinstall WKS software
0408011
Warning
Read serial port
failed
Workstation and peripheral
interface communicaiton
exception. Please restart the
system. If this issue persists,
please call for service.
Serial port of WKS abnormal
1. WKS serial port for UPS
may be abnormal.
2.Reinstall WKS software
0408012
Error
Set serial port
failed
Workstation and peripheral
interface communicaiton
exception. Please restart the
system. If this issue persists,
please call for service.
Serial port of WKS abnormal
1. WKS serial port for UPS
may be abnormal.
2.Reinstall WKS software
0408015
Warning
No battery in
UPS
No battery in UPS. Please keep
AC plugged in to avoid data
loss. If this issue persists.
Please call for service.
Summary of Contents for Brivo OEC 715
Page 2: ......
Page 19: ...Chapter1 Introduction and Safety...
Page 41: ...Introduction and Safety 23 46 54 20 18 26...
Page 55: ...Chapter2 System Overview...
Page 137: ...Chapter3 Installation...
Page 212: ...Chapter4 Calibration...
Page 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Page 284: ...Chapter5 Software...
Page 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Page 335: ...Chapter6 Troubleshooting...
Page 408: ...Chapter7 Replacement...
Page 418: ...Replacement 7 11 166...
Page 488: ...Chapter8 Periodic Maintenance...
Page 502: ...Periodic Maintenance 8 15...
Page 505: ...Chapter9 Technical Reference...
Page 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Page 526: ...11 Appendix System Schematics...