Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
7-64
Interconnect Cable Socket kit: (PN#: 5075972)
Interconnect Cable Socket Kit
(FRU PN: 5075972)
includes Interconnect cable and Interconnect
Cable socket. For existed IB system, FSE should replac
e the interconnect cable socket and
upgrade the interconnect cable(if Rev.13 or lower version).
Connector List:
Connector
To
Emergency Switch MF_AC_28V_N
Emergency Switch
Emergency Switch MF_AC_28V_L
J5
To exposure indicator
MBD MX7
To Mainframe Mother Board
AC220V_L
J4
AC220V_N
J4
PE
GND
N/A (Y/G Cable)
GND
Replacement procedure:
1.
Remove Mainframe back and front covers refer to section 7.
2.
Cut all the cable ties and unplug the connectors
FRU PN
5085708 Interconnect cable
5085709 Interconnect Cable socket
5075972
Interconnect
Cable Socket
Kit
Summary of Contents for Brivo OEC 715
Page 2: ......
Page 19: ...Chapter1 Introduction and Safety...
Page 41: ...Introduction and Safety 23 46 54 20 18 26...
Page 55: ...Chapter2 System Overview...
Page 137: ...Chapter3 Installation...
Page 212: ...Chapter4 Calibration...
Page 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Page 284: ...Chapter5 Software...
Page 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Page 335: ...Chapter6 Troubleshooting...
Page 408: ...Chapter7 Replacement...
Page 418: ...Replacement 7 11 166...
Page 488: ...Chapter8 Periodic Maintenance...
Page 502: ...Periodic Maintenance 8 15...
Page 505: ...Chapter9 Technical Reference...
Page 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Page 526: ...11 Appendix System Schematics...