Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
5-8
DAP Unit Config:
Field engineer can configure software DAP unit on this interface.
HLF level exposure setting:
Disable after High Level Exposure: Enable or disable “The system automatically Disable After High Level
Exposure” function. Some countries or regions, such as certain provincial locations in Australia, may
require enabling it.
Disable after No Exposure for 5 Minutes: Enable or disable “Disable after No Exposure for 5 Minutes”
function. Some countries or regions, such as certain provincial locations in Australia, may require
enabling it.
HLF Accessible Only in Auto ABS: Enable or disable “HLF Accessible Only in Auto ABS” function. Some
countries or regions, such as certain provincial locations in Australia, may require enabling it.
Automatic Termination at 20” in One Shot: Enable or disable “Automatic Termination at 20” in One Shot”
function. Some countries or regions, such as Australia or Switzerland, require enabling it.
Refer to the following table “Exposure Management Matrix” to set HLF scene setting.
Exposure Management Matrix
Countries /
Regions
Dose Group Option
HLF
Accessible
Only in
Auto ABS
Automatic
Termination
at 20” in
One Shot
Disable after
High Level
Exposure
Disable after
No Exposure
for 5
Minutes
Auto
Termination
at 9'30" in one
shot
Albania
Normal 88 and HLF 100 Disable
Enable
Disable
Disable
Enable
Algeria
Normal 88 and HLF 100 Disable
Enable
Disable
Disable
Enable
Angola
Normal 88 and HLF 100 Disable
Enable
Disable
Disable
Enable
Argentina
Normal 88 and HLF 100 Disable
Enable
Disable
Disable
Enable
Australia
Normal 50 and HLF 100 Enable
Enable
Enable
Enable
Enable
Austria
Normal 88 and HLF 100 Disable
Enable
Disable
Disable
Enable
Bahrain
Normal 88 and HLF 100 Disable
Enable
Disable
Disable
Enable
Bangladesh Normal 88 and HLF 100 Disable
Enable
Disable
Disable
Enable
Belarus
Normal 88 and HLF 100 Disable
Enable
Disable
Disable
Enable
Belgium
Normal 88 and HLF 100 Disable
Enable
Disable
Disable
Enable
Bolivia
Normal 88 and HLF 100 Disable
Enable
Disable
Disable
Enable
Bosnia
Normal 88 and HLF 100 Disable
Enable
Disable
Disable
Enable
Brazil
Normal 50 and HLF 100 Disable
Enable
Disable
Disable
Enable
Brunei
Normal 88 and HLF 100 Disable
Enable
Disable
Disable
Enable
Summary of Contents for Brivo OEC 715
Page 2: ......
Page 19: ...Chapter1 Introduction and Safety...
Page 41: ...Introduction and Safety 23 46 54 20 18 26...
Page 55: ...Chapter2 System Overview...
Page 137: ...Chapter3 Installation...
Page 212: ...Chapter4 Calibration...
Page 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Page 284: ...Chapter5 Software...
Page 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Page 335: ...Chapter6 Troubleshooting...
Page 408: ...Chapter7 Replacement...
Page 418: ...Replacement 7 11 166...
Page 488: ...Chapter8 Periodic Maintenance...
Page 502: ...Periodic Maintenance 8 15...
Page 505: ...Chapter9 Technical Reference...
Page 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...