Action
Step
Refill the syringe with distilled water and press gently to flush the flow path
through the flow restrictor. Be careful to avoid spillage if there is a salt
blockage creating back pressure. The flow restrictor itself will also create
some back pressure. Continue until the flow path is free.
6
Reconnect the pieces of tubing marked 2A, 1A1 and 1A2 and place the buffer
bottles on top of the system again.
7
If blockage is flushed away with water, stronger cleaning solutions can then
be used. Pump the cleaning solution through the pump in a standard manner
and out to the Injection valve waste oulet.
Note:
316
ÄKTA avant User Manual 29035184 AE
7 Maintenance
7.6 Maintenance when required
7.6.10 Clean pump flow restrictors