8.3
Troubleshooting: Monitors
In this section
•
UV monitor and UV detector
•
Conductivity monitor
•
pH monitor and pH valve
•
Pressure monitors
UV monitor U9-M and UV
detector unit
Possible cause and action
Problem
•
Communication problem
Contact Service.
•
The cable between the UV module and the ICU is
not connected
Remove the UV module and make sure that the cable
is connected.
•
Wrong Node ID
Check the module´s Node ID. If necessary, change the
Node ID. See Section 9.15 Check and change the Node
ID of a module, on page 515.
The UV module is not
found by the instru-
ment
ÄKTA avant User Manual 29035184 AE
363
8 Troubleshooting
8.3 Troubleshooting: Monitors