![GE AKTA avant User Manual Download Page 150](http://html.mh-extra.com/html/ge/akta-avant/akta-avant_user-manual_128005150.webp)
4.12.4
Connect external equipment to the I/O-box
Requirements on connected
equipment
The physical requirements for the connected equipment is described in the following
tables. All connected equipment must have a common grounding.
Analog input
Value
Parameter
2
Channels
±2000 mV
Range
1 MΩ
Input impedance
±(0.1% + 0.2 mV)
Accuracy
Analog output
Value
Parameter
2
Channels
±1000 mV
Range
100 kΩ
Input impedance
±(0.3% + 1 mV)
Accuracy
Digital input
Value
Parameter
4
Channels
TTL, open/closed circuit
Compatibility
Digital output
Value
Parameter
4
Channels
Open/closed circuit
Compatibility
150
ÄKTA avant User Manual 29035184 AE
4 Optional instrument modules
4.12 I/O-box
4.12.4 Connect external equipment to the I/O-box