Action
Step
Place the Cassette tray on the Tray support of the Frac drawer. Make sure
that the front of the tray (marked with the GE logo) faces the front of the
drawer.
6
Close the Frac drawer. Make sure that it snaps into closed position.
7
In the System Control module, on the View menu, click Fraction collection
content. In the Fraction collector content dialog box, check that the auto-
matic scanning procedures have been performed to show correct tube and
cassette positions.
8
In the System Control module, on the Systemmenu, click Settings.
9
Result: The System Settings dialog box opens.
In the System Settings dialog box:
10
•
Select Fraction collector:Fractionation settings.
•
For ÄKTA avant 25: In the Fractionation mode drop-down list, click Ac-
cumulator.
•
In the Fractionation order drop-down list, click Row-by-row.
•
Click OK.
Run the test
Follow the instructions described in Section 6.1 General performance test actions, on
page 240 to start and run the performance test.
ÄKTA avant User Manual 29035184 AE
247
6 Performance tests
6.3 Built-in Fraction collector test