9.2
Module specifications
Introduction
This section specifies the operating data of the components in ÄKTA avant. For general
data for the system see Section 9.1 System specifications, on page 422.
System pumps
Data
Parameter
Piston pump, metering type
Pump type
ÄKTA avant 25: 0.001 to 25 ml/min
Flow rate range (normal range)
ÄKTA avant 150: 0.01 to 150 ml/min
ÄKTA avant 25: 0.01 to 50 ml/min
Flow rate range (column packing flow)
ÄKTA avant 150: 0.01 to 300 ml/min
ÄKTA avant 25: 0 to 20 MPa (0 to 200 bar)
Pressure range
ÄKTA avant 150: 0 to 5 MPa (0 to 50 bar)
ÄKTA avant 25: 0.35 to 10 cP
Viscosity range
ÄKTA avant 150: 0.35 to 5 cP
ÄKTA avant 25:
Flow rate
specifications
•
Accuracy: ± 1.2%
•
Precision: RSD < 0.5%
(Conditions: 0.25 to 25 ml/min, < 3 MPa, 0.8 to 2 cP)
ÄKTA avant 150:
•
Accuracy: ± 1.5%
•
RSD < 0.5%
(Conditions: 1.0 to 150 ml/min, < 3 MPa, 0.8 to 2 cP)
424
ÄKTA avant User Manual 29035184 AE
9 Reference information
9.2 Module specifications
Summary of Contents for AKTA avant
Page 1: ...ÄKTA avant User Manual ...
Page 2: ...Page intentionally left blank ...
Page 536: ...Page intentionally left blank ...