Included phases
Principle
Predefined purification
method
After equilibration and sample application,
hydrophobic proteins adsorb to the column
ligand. After a wash to remove unbound
sample, elution is performed by generating
a gradient of a non-polar, organic solvent
such as acetonitrile. Finally, the column is
washed and re-equilibrated.
Reversed Phase Chro-
matography (RPC)
WARNING
Fraction collector. Do not fractionate flammable liquids in the
built-in fraction collector. When running RPC methods, collect
fractions through the outlet valve or the optional external Fraction
collector F9-R.
Predefined maintenance
methods
A number of predefined methods for preparation and cleaning are available. These
maintenance methods are used to prepare the system, clean the system, and to fill the
system with storage solution.
The following table describes the available predefined maintenance methods.
ÄKTA avant User Manual 29035184 AE
453
9 Reference information
9.6 Predefined methods and phases