Action
Step
In the Method Editor module,
5
•
click the Save the method button
or
•
on the File menu, click Save As
Result: The Save As dialog opens.
In the Save As dialog:
6
•
Select a target folder to enable the Save button.
•
Type a Name for the method.
•
Select a System from the list.
•
Click Save.
Result: The created method is saved in the selected folder.
Run a System CIP method
Follow the instructions to run a System CIP method.
Action
Step
In the Method Editor module, create a System CIP method according to the
previous instructions.
1
Connect by-pass tubing to all selected column positions and loop positions
if a loop valve is used.
2
Prepare cleaning solutions and immerse the selected inlet tubing in the so-
lutions.
Note:
Note that each phase uses one solution. All inlets selected in one phase should
be immersed in the same cleaning solution.
3
In the System Control module, select the created method and start the run.
4
For complete cleaning of the flow path, clean the manual injection port of
the Injection valve and the pH valve manually, see the instructions below.
5
ÄKTA avant User Manual 29035184 AE
293
7 Maintenance
7.6 Maintenance when required
7.6.2 Perform System CIP