Action
Step
In the Maintenance Manager dialog box, click New System Notification.
1
Result: The NewNotification field appears in the Maintenance Manager
dialog box.
In the NewNotification field:
2
•
Enter a name for the new notification.
•
Select a time interval after which the new notification will be issued.
•
If desired, write a message that will be shown for the maintenance noti-
fication.
Click Apply to save the changes and apply the notification settings.
3
Delete a user defined system
notification
To delete a user defined system notification, select the notification in the Maintenance
Manager dialog box and click Delete.
Module notifications are predefined and cannot be deleted. If desired, they can
be disabled.