Clean a pH electrode installed in
the pH valve
Follow the instructions below to clean a pH electrode installed in the pH valve. The cali-
bration function is used to switch the position of the pH valve. However, no calibration
is performed.
Action
Step
Open the System Control module. On the System menu, click Calibration.
1
Result: The Calibration dialog opens.
Set the pH monitor as the monitor to calibrate by clicking pH on the Monitor
to calibrate menu.
2
Click Prepare for Calibration.
3
Result: The pH valve switches to the calibration position.
ÄKTA avant User Manual 29035184 AE
309
7 Maintenance
7.6 Maintenance when required
7.6.7 Clean the pH electrode