3.7
Outlet valve
Function of the Outlet valve
The Outlet valve is used to direct the flow to the fraction collector, to an outlet port, or
to waste. The Outlet valve enables collection of fractions. Fractions can be collected by
using a fraction collector connected to the Outlet valve, or by using the 10 outlet ports.
Usage of the outlet ports can be convenient when, for example, a limited number of
large fractions is to be collected. If no fraction collection is needed, the Outlet valve can
be set to direct the flow to waste.
Location and illustration of
Outlet valve
The following illustration shows the location of the Outlet valve and also an enlargement
of the valve. In ÄKTA avant 25, the Outlet valve is labelled V9-O. In ÄKTA avant 150, the
Outlet valve is labelled V9H-O.
76
ÄKTA avant User Manual 29035184 AE