Text Instruction mode
In most cases methods can be edited using the Phase Properties pane in the Method
Editor module. However, modules without a recommended position are not supported
by Phase Properties and have to be edited in the Text Instruction mode.
The following modules require that the method is created using the Text Instruction
mode:
•
Inlet valve IX
It is necessary to use instructions for both Inlet valve IX (V9-IX or V9H-IX)
and the other inlet valve.
Note:
•
Versatile valve V9-V or V9H-V
•
I/O-box E9
•
UV monitor U9-L, 2nd
The UV monitor U9-L, 2nd is auto zeroed at the same time as UV monitor
U9-M in predefined methods. All other UV monitor U9-L, 2nd instructions
need to be edited in the Text Instruction mode.
Note:
ÄKTA avant User Manual 29035184 AE
229
5 Operation
5.9 Create a method and perform a run
5.9.1 Create a method