8.5
Troubleshooting: Fraction collector
In this section
•
Built-in Fraction collector
•
Fraction collector F9-R
Built-in fraction collector
Only ÄKTA avant 25 has DropSync sensors.
Note:
Possible cause and action
Problem
•
The cable between the Fraction collector and the
ICU is not connected
Generate a System error report and contact Service.
•
A fuse in the instrument ICU is broken
The ICU needs to be changed. Generate a System error
report and contact Service.
•
Wrong Node ID
•
Check the module's Node ID. If necessary, change the
Node ID. See Section 9.15 Check and change the Node
ID of a module, on page 515.
The Fraction collector
cannot be found by the
instrument
•
Obstruction inside the Fraction collector
Switch off the instrument and check for obstruction
inside the Fraction collector. Try to move the Frac arm
by hand. Switch on the instrument. If this error is recur-
rent, generate a System error report and contact Ser-
vice.
Fraction collector arm
is blocked or internal
fault in the Fraction
collector
•
Salt crystals or protein residuals block the accumu-
lator
Restart the instrument and perform an accumulator
wash.
•
Mechanical error
If this error is recurrent, generate a System error report
and contact Service.
The accumulator is
jammed or there is an
internal error in the in-
strument
384
ÄKTA avant User Manual 29035184 AE
8 Troubleshooting
8.5 Troubleshooting: Fraction collector