7.8.5
Replace UV monitor U9-L flow cell
Maintenance interval
Replace the UV flow cell when it is desired to use a flow cell with a different path length,
or if the cell is damaged.
Required material
•
UV flow cell
Replace the flow cell
CAUTION
Hazardous chemicals or biological agents in UV flow cell. Make
sure that the entire flow cell has been flushed thoroughly with
bacteriostatic solution (e.g., NaOH) and distilled water, before service
and maintenance.
Follow the instruction to replace the UV flow cell.
Action
Step
Switch off the instrument.
1
Disconnect the tubing from the UV flow cell.
2
ÄKTA avant User Manual 29035184 AE
345
7 Maintenance
7.8 Replacement procedures
7.8.5 Replace UV monitor U9-L flow cell