13-35
Chapter 13 Menu Items
Display Configuration of the Home Display
NOTE
If [OFF] is set for "Center Split" under "Menu>Common Setup>Display Config.>All
Beds>Layout Change", numeric data cannot be displayed in the waveform area.
If [Name] or [Bed Name] is set for "Waveform Area" under "Menu>Common
Setup>Display Config.>Detail Setup>, maximum of 2 numeric data can be displayed.
If [Patient Name] is set for [Waveform] under "Menu>Common Setup>Display
Config.>Each Bed", numeric data cannot be displayed in the waveform area.
For the bed with one (1) waveform display, numeric data cannot be displayed in the
waveform area.
The following numeric data cannot be displayed in the waveform area.
"ST-A" , "ST-B" , "ST-C" , "T1, T2" , "T3, T4" , "T5, T6" , "T7, T8" , "SvO
2
, CO" , "RR,
CO
2
, Agent, O
2
, N
2
O" , "CO
2
, Agent, O
2
, N
2
O" , "RR, Agent, O
2
, N
2
O" , "Agent, O
2
,
N
2
O" , "Agent, N
2
O" , "SPIRO" , "GAS, SPIRO" , "VENT" ,
The numeric data on the waveform area will be displayed with the following priority.
High Priority: Common Setup>Display Config.>Each Bed>Numeric Data
Low Priority: Common Setup>Display Config.>Detail Setup>Numeric Data>Display Priority
The highest priority is the numeric data set at the top left on the "Each Bed" setting, and the lowest
priority is the numeric data set at the bottom on the "Display Priority" setting.
Exiting the Display Configuration Setup
To end the display configuration setup, follow the procedure below.
1
Press the
or
key or any user keys other than [Home] to change the screen.
A confirmation message will be displayed.
2
To register the settings, press the [Register] key.
[Cancel] : The settings will be cancelled.
[Prev. Disp]: The display will return to the display
configuration menu.
Setup Item
Priority
"Each Bed" setting: Top Left
High
"Each Bed" setting: Top Right
↑
"Each Bed" setting: Bottom Left
"Each Bed" setting: Bottom Right
"Display Priority" setting: Top
↓
"Display Priority" setting: Bottom
Low
Summary of Contents for DYASCOPE 8000 Series
Page 1: ......
Page 2: ......
Page 12: ...Contents ...
Page 20: ...Contents ...
Page 44: ...1 2 Chapter 1 General Description Contents ...
Page 50: ...2 2 Chapter 2 Name of Parts and Their Functions Contents ...
Page 54: ...2 4 Chapter 2 Name of Parts and Their Functions ...
Page 56: ...3 2 Chapter 3 Description of the Display Contents ...
Page 70: ...4 2 Chapter 4 Basic Operation Contents ...
Page 82: ...5 2 Chapter 5 Preparation Contents ...
Page 88: ...5 6 Chapter 5 Preparation Nurse Call Daily Check ...
Page 90: ...6 2 Chapter 6 Admit Discharge Contents ...
Page 110: ...7 2 Chapter 7 Alarm Function Contents ...
Page 138: ...8 2 Chapter 8 Parameter Setup Contents ...
Page 164: ...8 26 Chapter 8 Parameter Setup Parameter ON OFF ...
Page 166: ...9 2 Chapter 9 Data Review Contents ...
Page 190: ...10 2 Chapter 10 Waveform Review Contents ...
Page 206: ...11 2 Chapter 11 Calculation Contents ...
Page 212: ...12 2 Chapter 12 Printing Contents ...
Page 230: ...13 2 Chapter 13 Menu Items Contents ...
Page 270: ...13 40 Chapter 13 Menu Items Nurse Team Setup ...
Page 272: ...14 2 Chapter 14 Troubleshooting Contents ...
Page 296: ...14 24 Chapter 14 Troubleshooting Troubleshooting ...
Page 298: ...15 2 Chapter 15 Setup Item Default Value Contents ...
Page 320: ...15 22 Chapter 15 Setup Item Default Value Menu Central Monitor Display ...
Page 322: ...16 2 Chapter 16 Accessories Contents ...
Page 326: ...16 4 Chapter 16 Accessories Optional Accessories ...
Page 328: ...17 2 Chapter 17 Specification Contents ...
Page 342: ...17 14 Chapter 17 Specification Technical Information ...
Page 347: ......
Page 348: ......
Page 349: ......