12-4
Chapter 12 Printing
Printing Condition/Output Destination Setup
[OFF]: Printing will not start at alarm occurrence.
3
Alarm Factor Selection
Select the alarm factor for alarm printing. When selected, the key will be displayed in blue.
The alarm OFF mark
will be displayed inside the key for the parameter in alarm OFF condition.
[Select All Arrhythmia]: All arrhythmia factors will be selected.
[All ON]: All alarm factors will be selected.
[All OFF]: All selections for the alarm factor will be cancelled.
4
Output Waveform Selection
Maximum of 3 waveforms can be selected for printing.
[Alarm]: Prints the waveform of the alarm factor.
(
"Alarm Factor and Printed Waveform" P12-4)
5
Print Duration
[12sec.]/[24sec.]: Printing will automatically stop after 12 seconds or 24 seconds.
The delay time differs depending on the print duration.
6
All Beds
Applies the same setting for alarm printing to all beds.
Alarm Factor and Printed Waveform
Print Duration
Delay Time
Adult
Child
Neonate
Numeric Data Alarm
Arrhythmia Alarm
12 sec.
12 sec.
12 sec.
8 sec.
12 sec.
8 sec. for the multigas unit alarm
24 sec.
16 sec.
16 sec.
16 sec.
16 sec.
Alarm Factor
Printed Waveform
Alarm Factor
Printed Waveform
Numeric Data
Numeric Data
HR
ECG1, ECG2
SpHb
SpO
2
ST1, ST2
ECG1, ECG2
SpHb-2
SpO
2
-2
ST(I) to ST(V6)
ECG1, ECG2
InspO
2
, ExpO
2
O
2
BP1 to 8, CVP
BP1 to 8
InspN
2
O, ExpN
2
O
CO
2
NIBP
-
InspAGT, ExpAGT
AGENT
SpO
2
SpO
2
InspAGT2, ExpAGT2
AGENT
SpO
2
-2
SpO
2
-2
MAC
AGENT
PR
SpO
2
Arrhythmia
PR-2
SpO
2
-2
Asystole
ECG1, ECG2
RR
RESP (Other than CO
2
source)
VF
ECG1, ECG2
CO
2
(CO
2
Source)
VT
ECG1, ECG2
APNEA
RESP (Other than CO
2
source)
Slow VT
ECG1, ECG2
CO
2
(CO
2
Source)
Run
ECG1, ECG2
EtCO
2
, InspCO
2
CO
2
Couplet
ECG1, ECG2
Summary of Contents for DYASCOPE 8000 Series
Page 1: ......
Page 2: ......
Page 12: ...Contents ...
Page 20: ...Contents ...
Page 44: ...1 2 Chapter 1 General Description Contents ...
Page 50: ...2 2 Chapter 2 Name of Parts and Their Functions Contents ...
Page 54: ...2 4 Chapter 2 Name of Parts and Their Functions ...
Page 56: ...3 2 Chapter 3 Description of the Display Contents ...
Page 70: ...4 2 Chapter 4 Basic Operation Contents ...
Page 82: ...5 2 Chapter 5 Preparation Contents ...
Page 88: ...5 6 Chapter 5 Preparation Nurse Call Daily Check ...
Page 90: ...6 2 Chapter 6 Admit Discharge Contents ...
Page 110: ...7 2 Chapter 7 Alarm Function Contents ...
Page 138: ...8 2 Chapter 8 Parameter Setup Contents ...
Page 164: ...8 26 Chapter 8 Parameter Setup Parameter ON OFF ...
Page 166: ...9 2 Chapter 9 Data Review Contents ...
Page 190: ...10 2 Chapter 10 Waveform Review Contents ...
Page 206: ...11 2 Chapter 11 Calculation Contents ...
Page 212: ...12 2 Chapter 12 Printing Contents ...
Page 230: ...13 2 Chapter 13 Menu Items Contents ...
Page 270: ...13 40 Chapter 13 Menu Items Nurse Team Setup ...
Page 272: ...14 2 Chapter 14 Troubleshooting Contents ...
Page 296: ...14 24 Chapter 14 Troubleshooting Troubleshooting ...
Page 298: ...15 2 Chapter 15 Setup Item Default Value Contents ...
Page 320: ...15 22 Chapter 15 Setup Item Default Value Menu Central Monitor Display ...
Page 322: ...16 2 Chapter 16 Accessories Contents ...
Page 326: ...16 4 Chapter 16 Accessories Optional Accessories ...
Page 328: ...17 2 Chapter 17 Specification Contents ...
Page 342: ...17 14 Chapter 17 Specification Technical Information ...
Page 347: ......
Page 348: ......
Page 349: ......