13-11
Chapter 13 Menu Items
Bed Transfer/Bed Exchange
By performing central monitor communication setup, bed transfer/exchange among several central monitors can be
performed through the TCP/IP network.
CAUTION
When a bed transfer procedure is performed, all setup data for the new bed will be updated.
The data for the DS-LAN bed, LW-T bed, and the same data monitored on other central
monitor will be initialized.
If the bed transfer/exchange is performed for the DS-LAN bed, the GAS alarm settings will
be backed up or initialized depending on the settings for "Backup at Discharge" on the
bedside monitor.
When using the EMR link function, bed transfer/exchange cannot be performed.
However, bed transfer/exchange can be performed when the EMR link is offline.
NOTE
Perform settings for the "Central Monitor Comm." in advance. (Initial Settings > External
Device > Network)
(
Maintenance Manual "Central Monitor Communication" P2-23)
1
Press the [Menu], [Bed Transfer] ("Function") keys.
The "Bed Transfer" screen will be displayed.
Details
Bed Transfer
Bed Exchange
Original
New
Original
New
Patient Information
Admit Settings
Discharge *
1
Transfer
Transfer
Transfer
Alarm
Alarm Settings
Discharge *
2
Transfer
Transfer
Transfer
Print Settings
Print Settings
Discharge *
3
Transfer*
7
Transfer*
7
Transfer*
7
Review Data
Graphic Trend
Discharge *
1
Transfer*
6
Transfer*
6
Transfer*
6
Tabular Trend
Full Disclosure Waveform
Recall
ST
Delete or
Transfer*
4
Delete or
Transfer*
4
Delete or
Transfer*
4
12-Lead Analysis
Hemodynamics
Nurse Call Setup
Nurse Call Settings
Discharge *
2
No Change*
5
No Change*
5
No Change*
5
Initial Settings
When PHS nurse call system is used
Bed Name Settings
Initialize
Initialize
Initialize
Initialize
*
1
: The data will be initialized or deleted.
*
2
: The initial settings at admittance will be applied.
*
3
: The settings will be backed up.
*
4
: The data will be deleted if between different central monitors, and the data will be transferred if within the same central monitor.
*
5
: The settings will not change and the same settings will be applied after bed transfer/exchange.
*
6
: If the bed transfer is performed within the same central monitor, the data will be transferred, and if the bed transfer is performed
between the different central monitors, the original data can be monitored via TCP/IP network.
*
7
: If the bed transfer is performed within the same central monitor, the settings will be transferred, and if the bed transfer is
performed between the different central monitors, the settings will not transfer.
Summary of Contents for DYASCOPE 8000 Series
Page 1: ......
Page 2: ......
Page 12: ...Contents ...
Page 20: ...Contents ...
Page 44: ...1 2 Chapter 1 General Description Contents ...
Page 50: ...2 2 Chapter 2 Name of Parts and Their Functions Contents ...
Page 54: ...2 4 Chapter 2 Name of Parts and Their Functions ...
Page 56: ...3 2 Chapter 3 Description of the Display Contents ...
Page 70: ...4 2 Chapter 4 Basic Operation Contents ...
Page 82: ...5 2 Chapter 5 Preparation Contents ...
Page 88: ...5 6 Chapter 5 Preparation Nurse Call Daily Check ...
Page 90: ...6 2 Chapter 6 Admit Discharge Contents ...
Page 110: ...7 2 Chapter 7 Alarm Function Contents ...
Page 138: ...8 2 Chapter 8 Parameter Setup Contents ...
Page 164: ...8 26 Chapter 8 Parameter Setup Parameter ON OFF ...
Page 166: ...9 2 Chapter 9 Data Review Contents ...
Page 190: ...10 2 Chapter 10 Waveform Review Contents ...
Page 206: ...11 2 Chapter 11 Calculation Contents ...
Page 212: ...12 2 Chapter 12 Printing Contents ...
Page 230: ...13 2 Chapter 13 Menu Items Contents ...
Page 270: ...13 40 Chapter 13 Menu Items Nurse Team Setup ...
Page 272: ...14 2 Chapter 14 Troubleshooting Contents ...
Page 296: ...14 24 Chapter 14 Troubleshooting Troubleshooting ...
Page 298: ...15 2 Chapter 15 Setup Item Default Value Contents ...
Page 320: ...15 22 Chapter 15 Setup Item Default Value Menu Central Monitor Display ...
Page 322: ...16 2 Chapter 16 Accessories Contents ...
Page 326: ...16 4 Chapter 16 Accessories Optional Accessories ...
Page 328: ...17 2 Chapter 17 Specification Contents ...
Page 342: ...17 14 Chapter 17 Specification Technical Information ...
Page 347: ......
Page 348: ......
Page 349: ......