8-5
Chapter 8 Parameter Setup
ECG
below -20 mm / -2.0 mV is set, the lower alarm will turn OFF.
Size / Lead
Adjustment of Waveform Size and Baseline Position
Adjust the waveform size and baseline position.
CAUTION
Automatic size of the ECG is effective only at the time the [Auto] key is pressed. This does
not continually adjust the size.
The waveform size cannot be set if the waveform is not displayed. Change the display
configuration as necessary.
(
"Numeric Data/Waveform" P13-4)
1
Press the key for "ECG1" to "ECG12".
The "Size" screen will be displayed.
When the display layout is "12-Lead", the waveform size can be set
differently for limb leads and chest leads.
2
Select the waveform size for displaying and printing.
[Auto]: ECG amplitude will be automatically adjusted to 10mm.
The automatic adjustment is effective only when the [Auto] key is
pressed.
3
Use the
/
keys to adjust the baseline position.
If the waveform is difficult to see due to ECG amplitude, set the baseline position to 0mV. The baseline position
for the waveform display and printing will be adjusted. When the display layout is set to "12-Lead", the baseline
position cannot be changed.
Waveform Size
x1/4
x1/2
x1
x2
x4
Voltage (10mm)
4mV
2mV
1mV
500μV
250μV
3
2
Summary of Contents for DYASCOPE 8000 Series
Page 1: ......
Page 2: ......
Page 12: ...Contents ...
Page 20: ...Contents ...
Page 44: ...1 2 Chapter 1 General Description Contents ...
Page 50: ...2 2 Chapter 2 Name of Parts and Their Functions Contents ...
Page 54: ...2 4 Chapter 2 Name of Parts and Their Functions ...
Page 56: ...3 2 Chapter 3 Description of the Display Contents ...
Page 70: ...4 2 Chapter 4 Basic Operation Contents ...
Page 82: ...5 2 Chapter 5 Preparation Contents ...
Page 88: ...5 6 Chapter 5 Preparation Nurse Call Daily Check ...
Page 90: ...6 2 Chapter 6 Admit Discharge Contents ...
Page 110: ...7 2 Chapter 7 Alarm Function Contents ...
Page 138: ...8 2 Chapter 8 Parameter Setup Contents ...
Page 164: ...8 26 Chapter 8 Parameter Setup Parameter ON OFF ...
Page 166: ...9 2 Chapter 9 Data Review Contents ...
Page 190: ...10 2 Chapter 10 Waveform Review Contents ...
Page 206: ...11 2 Chapter 11 Calculation Contents ...
Page 212: ...12 2 Chapter 12 Printing Contents ...
Page 230: ...13 2 Chapter 13 Menu Items Contents ...
Page 270: ...13 40 Chapter 13 Menu Items Nurse Team Setup ...
Page 272: ...14 2 Chapter 14 Troubleshooting Contents ...
Page 296: ...14 24 Chapter 14 Troubleshooting Troubleshooting ...
Page 298: ...15 2 Chapter 15 Setup Item Default Value Contents ...
Page 320: ...15 22 Chapter 15 Setup Item Default Value Menu Central Monitor Display ...
Page 322: ...16 2 Chapter 16 Accessories Contents ...
Page 326: ...16 4 Chapter 16 Accessories Optional Accessories ...
Page 328: ...17 2 Chapter 17 Specification Contents ...
Page 342: ...17 14 Chapter 17 Specification Technical Information ...
Page 347: ......
Page 348: ......
Page 349: ......