14-1
Chapter 14 Troubleshooting
Message List
Chapter 14 Troubleshooting
Message List
In this section, the displayed messages on this equipment are described.
For the vital alarm message, there are numeric data alarm and arrhythmia alarm, and the delay time are as follows.
Numeric Data Alarm: Adult/Child: 5 sec., Neonate: none
Arrhythmia Alarm: Adult/Child/Neonate: none
CAUTION
The alarm level can be changed on the "Alarm Level" under the "Initial Settings".
The following table shows the alarm level selection for each parameter.
There are 5 alarm levels.
Numeric Data Alarm
Ex.) HR alarm level : "S H M x x"
S:
Top Priority (Top Priority Alarm)
H:
High Priority (Urgent Alarm)
M:
Medium Priority (Cautionary Alarm)
L:
Low Priority (Status Alarm)
N:
Notification (Notification Alarm)
The alarm level that cannot be set will be indicated by "x".
For the HR Alarm, the alarm level of S (Top) / H (High) / M (Medium) can be set, and L (Low) / N (Notification) cannot be set.
Parameter
Message
Default
Level
Alarm Level
Selection
Note
-
<Alarm Susp: xxx s>
N
x x x x N
"xxx s" indicates remaining time.
ECG
<Upper HR>
H
S H M x x
<Lower HR>
ST
<Upper ST1>
M
x H M x x
<Lower ST1>
<Upper ST2>
M
x H M x x
<Lower ST2>
12-Lead ST
<Upper ST (xxx)>
M
x H M x x
(xxx) indicates lead I to V6.
<Lower ST (xxx)>
BP1 to BP8
<Upper ###>
H/M
x H M x x
### indicates BP label.
Default Alarm Level
BP1: H, BP2 to BP8: M
<Lower ###>
<Upper PR_IBP>
H
x H M x x
<Lower PR_IBP>
Summary of Contents for DYASCOPE 8000 Series
Page 1: ......
Page 2: ......
Page 12: ...Contents ...
Page 20: ...Contents ...
Page 44: ...1 2 Chapter 1 General Description Contents ...
Page 50: ...2 2 Chapter 2 Name of Parts and Their Functions Contents ...
Page 54: ...2 4 Chapter 2 Name of Parts and Their Functions ...
Page 56: ...3 2 Chapter 3 Description of the Display Contents ...
Page 70: ...4 2 Chapter 4 Basic Operation Contents ...
Page 82: ...5 2 Chapter 5 Preparation Contents ...
Page 88: ...5 6 Chapter 5 Preparation Nurse Call Daily Check ...
Page 90: ...6 2 Chapter 6 Admit Discharge Contents ...
Page 110: ...7 2 Chapter 7 Alarm Function Contents ...
Page 138: ...8 2 Chapter 8 Parameter Setup Contents ...
Page 164: ...8 26 Chapter 8 Parameter Setup Parameter ON OFF ...
Page 166: ...9 2 Chapter 9 Data Review Contents ...
Page 190: ...10 2 Chapter 10 Waveform Review Contents ...
Page 206: ...11 2 Chapter 11 Calculation Contents ...
Page 212: ...12 2 Chapter 12 Printing Contents ...
Page 230: ...13 2 Chapter 13 Menu Items Contents ...
Page 270: ...13 40 Chapter 13 Menu Items Nurse Team Setup ...
Page 272: ...14 2 Chapter 14 Troubleshooting Contents ...
Page 296: ...14 24 Chapter 14 Troubleshooting Troubleshooting ...
Page 298: ...15 2 Chapter 15 Setup Item Default Value Contents ...
Page 320: ...15 22 Chapter 15 Setup Item Default Value Menu Central Monitor Display ...
Page 322: ...16 2 Chapter 16 Accessories Contents ...
Page 326: ...16 4 Chapter 16 Accessories Optional Accessories ...
Page 328: ...17 2 Chapter 17 Specification Contents ...
Page 342: ...17 14 Chapter 17 Specification Technical Information ...
Page 347: ......
Page 348: ......
Page 349: ......