Chapter 14 Troubleshooting
Message List
NOTE
<NIBP meas. failed> alarm will be canceled when [Alarm Silence] key is pressed. Pay
attention not to cancel the important alarm.
System Status
<Failed to upload.>
N
Data Transfer: Failed to upload transferring data from the bedside
monitor to DS-8900.
<Upload Standby>
N
Data Transfer: Uploading of other bed is in progress.
<Transmitting Data>
N
Data Transfer: Transmitting data from the DS-8900 to data server.
<Chk Data Transfer>
N
Data Transfer: Failed to transmit data from the DS-8900 to data server.
Message
Level
Note
Message
Level
Details
<Central ID is duplicated.>
M
Central ID is duplicated with other central monitor.
<Check PHS comm.>
M
Communication error with the PHS nurse call system.
<Check PHS call target>
N
All the PHS nurse call targets are not functioning.
<Check PHS monitor ID>
N
Monitor ID is not set for the PHS nurse call system.
<Updating PHS room info.>
M
Receiving room information from the nurse call system controller.
<Check TCON Comm.>
N
Communication error with the TCON system.
<TCON Interference>
N
Noise is interfering.
<Chk TCON Reception>
N
When TCON "C2" is set, communication with the TCON administrator (C1)
fails.
<Check extended memory>
M
Error on the extended memory.
<Check PC/CF Card>
N
CF card error, or the card cannot be identified.
<Check Data Server Comm.>
N
When a data server is used, communication with the data server fails.
<Check EMR comm.>
N
When EMR link function is used, communication with the patient data server
fails.
<Check Patient Server Comm.>
N
When "Search ID" is used, communication with the patient data server fails.
<Check SNTP Comm.>
N
When a SNTP server is used, communication with the SNTP server fails.
<Check multimonitor conn.>
N
Serial communication with the extended display unit fails.
<DS-8900 Check Unit>
L
The control part of the main unit is malfunctioning.
<DS-8900 Failure>
H
The control part of the main unit is inoperative.
<Speaker Failure>
H
The speaker is inoperative.
<Check Backup Battery>
N
Backup battery is depleted.
<EMR Offline>
N
When EMR link function is used, EMR offline is set.
Summary of Contents for DYASCOPE 8000 Series
Page 1: ......
Page 2: ......
Page 12: ...Contents ...
Page 20: ...Contents ...
Page 44: ...1 2 Chapter 1 General Description Contents ...
Page 50: ...2 2 Chapter 2 Name of Parts and Their Functions Contents ...
Page 54: ...2 4 Chapter 2 Name of Parts and Their Functions ...
Page 56: ...3 2 Chapter 3 Description of the Display Contents ...
Page 70: ...4 2 Chapter 4 Basic Operation Contents ...
Page 82: ...5 2 Chapter 5 Preparation Contents ...
Page 88: ...5 6 Chapter 5 Preparation Nurse Call Daily Check ...
Page 90: ...6 2 Chapter 6 Admit Discharge Contents ...
Page 110: ...7 2 Chapter 7 Alarm Function Contents ...
Page 138: ...8 2 Chapter 8 Parameter Setup Contents ...
Page 164: ...8 26 Chapter 8 Parameter Setup Parameter ON OFF ...
Page 166: ...9 2 Chapter 9 Data Review Contents ...
Page 190: ...10 2 Chapter 10 Waveform Review Contents ...
Page 206: ...11 2 Chapter 11 Calculation Contents ...
Page 212: ...12 2 Chapter 12 Printing Contents ...
Page 230: ...13 2 Chapter 13 Menu Items Contents ...
Page 270: ...13 40 Chapter 13 Menu Items Nurse Team Setup ...
Page 272: ...14 2 Chapter 14 Troubleshooting Contents ...
Page 296: ...14 24 Chapter 14 Troubleshooting Troubleshooting ...
Page 298: ...15 2 Chapter 15 Setup Item Default Value Contents ...
Page 320: ...15 22 Chapter 15 Setup Item Default Value Menu Central Monitor Display ...
Page 322: ...16 2 Chapter 16 Accessories Contents ...
Page 326: ...16 4 Chapter 16 Accessories Optional Accessories ...
Page 328: ...17 2 Chapter 17 Specification Contents ...
Page 342: ...17 14 Chapter 17 Specification Technical Information ...
Page 347: ......
Page 348: ......
Page 349: ......