Chapter 14 Troubleshooting
Troubleshooting
PHS Nurse Call System
EMR Link Function
Situation
Cause
Solution
The <Check PHS comm.> message is
displayed.
Or, <Failed> is displayed as a result of
Nurse Call Daily Check.
Communication error with nurse call
system.
Check the connection with the nurse call
system.
(
Maintenance Manual "Connecting the
Nurse Call System" P5-3)
On the "Nurse Call Daily Check" screen,
the check key ( [1] / [2] / [3] ) remains to
be lit in blue and other check keys do not
function.
The check key lit in blue is in process of
nurse call.
The key will remain blue until the nurse call
connection is verified at the call target (base
station, PHS, etc.). Verify the connection at the
call target.
The <Check PHS call target> message is
displayed.
All the PHS nurse call targets (base
station, PHS, etc.) are not functioning.
Check the power and connection of the nurse
call targets (base station, PHS, etc.).
The <Check PHS monitor ID> message is
displayed.
For the "Detail Setup" of "Nurse Call"
(Initial Settings > External Devices >
Serial Comm.), "Monitor ID" is set to [0].
"Monitor ID" should be set in the range from 1
to 99, and make sure that it does not duplicate
with other central monitors.
(
Maintenance Manual "Nurse Call Detail
Setup" P5-4)
Situation
Cause
Solution
The <Check EMR comm.>
message is displayed. The <Check
Data Server Comm.> message is
displayed.
There is a communication failure between
the DS-8900 system and the patient data
server.
Check the communication status of the DS-8900
system, patient data server, and EMR machine.
The network setup is incorrect.
Set the correct IP address and port number on the
network setup menu (Initial Settings > External
Device > Network).
(
Maintenance Manual "Patient Data Server"
P2-24)
The patient data server system has gone
down.
Check if the patient data server is properly
operating. If not, refer to the operation manual of
the patient data server.
The connection cable is disconnected.
Verify that the DS-8900 system, patient data server,
EMR machine is connected properly. If not, connect
them properly.
The protocol of data server setting on this
equipment does not correspond with the
setting on the data server.
Check if protocol of data server setting on this
equipment corresponds with the setting on the data
server. If not, set it correctly.
Summary of Contents for DYASCOPE 8000 Series
Page 1: ......
Page 2: ......
Page 12: ...Contents ...
Page 20: ...Contents ...
Page 44: ...1 2 Chapter 1 General Description Contents ...
Page 50: ...2 2 Chapter 2 Name of Parts and Their Functions Contents ...
Page 54: ...2 4 Chapter 2 Name of Parts and Their Functions ...
Page 56: ...3 2 Chapter 3 Description of the Display Contents ...
Page 70: ...4 2 Chapter 4 Basic Operation Contents ...
Page 82: ...5 2 Chapter 5 Preparation Contents ...
Page 88: ...5 6 Chapter 5 Preparation Nurse Call Daily Check ...
Page 90: ...6 2 Chapter 6 Admit Discharge Contents ...
Page 110: ...7 2 Chapter 7 Alarm Function Contents ...
Page 138: ...8 2 Chapter 8 Parameter Setup Contents ...
Page 164: ...8 26 Chapter 8 Parameter Setup Parameter ON OFF ...
Page 166: ...9 2 Chapter 9 Data Review Contents ...
Page 190: ...10 2 Chapter 10 Waveform Review Contents ...
Page 206: ...11 2 Chapter 11 Calculation Contents ...
Page 212: ...12 2 Chapter 12 Printing Contents ...
Page 230: ...13 2 Chapter 13 Menu Items Contents ...
Page 270: ...13 40 Chapter 13 Menu Items Nurse Team Setup ...
Page 272: ...14 2 Chapter 14 Troubleshooting Contents ...
Page 296: ...14 24 Chapter 14 Troubleshooting Troubleshooting ...
Page 298: ...15 2 Chapter 15 Setup Item Default Value Contents ...
Page 320: ...15 22 Chapter 15 Setup Item Default Value Menu Central Monitor Display ...
Page 322: ...16 2 Chapter 16 Accessories Contents ...
Page 326: ...16 4 Chapter 16 Accessories Optional Accessories ...
Page 328: ...17 2 Chapter 17 Specification Contents ...
Page 342: ...17 14 Chapter 17 Specification Technical Information ...
Page 347: ......
Page 348: ......
Page 349: ......