13-23
Chapter 13 Menu Items
Full Disclosure Waveform Setup
Select from [5sec.]/ [10sec.]/ [15sec.]/ [20sec.]/ [30sec.]/ [None]/ [None (Noise OFF)].
For HR, Tachy, Brady, Ext Tachy, Ext Brady, and the custom factor including these parameters, [None
(Noise OFF)] will be displayed.
[None], [Noise OFF]: Alarm will be notified to the nurse call system at alarm generation without any
delay.
REFERENCE
When [None (Noise OFF)] is selected, noise detection will be performed before nurse
call notification. If detected as noise, the alarm will not be notified to the nurse call
system.
For "Arrhythmia Alarm", "APNEA", "NIBP", "Too Far", alarm duration before
notification cannot be set. The alarm will be notified to the nurse call system at alarm
generation without any delay.
4
To set the same setting for all beds, press the [Setup] key for "All Beds".
Full Disclosure Waveform Setup
REFERENCE
The parameters to save as full disclosure waveform can be set. Refer to the following
chapter.
(
"To Select the Waveform to Store" P10-7)
Data Server Output Waveform Setup
The beds and waveforms to output to the data server can be set. There are following 2 types of settings.
Displayed Beds: Maximum of 16 beds which are displayed on the home display.
Registered Beds: Maximum of 32 beds which are previously registered.
1
Press the [Menu], [Data Server Waveform] ("Each Bed") keys.
Summary of Contents for DYASCOPE 8000 Series
Page 1: ......
Page 2: ......
Page 12: ...Contents ...
Page 20: ...Contents ...
Page 44: ...1 2 Chapter 1 General Description Contents ...
Page 50: ...2 2 Chapter 2 Name of Parts and Their Functions Contents ...
Page 54: ...2 4 Chapter 2 Name of Parts and Their Functions ...
Page 56: ...3 2 Chapter 3 Description of the Display Contents ...
Page 70: ...4 2 Chapter 4 Basic Operation Contents ...
Page 82: ...5 2 Chapter 5 Preparation Contents ...
Page 88: ...5 6 Chapter 5 Preparation Nurse Call Daily Check ...
Page 90: ...6 2 Chapter 6 Admit Discharge Contents ...
Page 110: ...7 2 Chapter 7 Alarm Function Contents ...
Page 138: ...8 2 Chapter 8 Parameter Setup Contents ...
Page 164: ...8 26 Chapter 8 Parameter Setup Parameter ON OFF ...
Page 166: ...9 2 Chapter 9 Data Review Contents ...
Page 190: ...10 2 Chapter 10 Waveform Review Contents ...
Page 206: ...11 2 Chapter 11 Calculation Contents ...
Page 212: ...12 2 Chapter 12 Printing Contents ...
Page 230: ...13 2 Chapter 13 Menu Items Contents ...
Page 270: ...13 40 Chapter 13 Menu Items Nurse Team Setup ...
Page 272: ...14 2 Chapter 14 Troubleshooting Contents ...
Page 296: ...14 24 Chapter 14 Troubleshooting Troubleshooting ...
Page 298: ...15 2 Chapter 15 Setup Item Default Value Contents ...
Page 320: ...15 22 Chapter 15 Setup Item Default Value Menu Central Monitor Display ...
Page 322: ...16 2 Chapter 16 Accessories Contents ...
Page 326: ...16 4 Chapter 16 Accessories Optional Accessories ...
Page 328: ...17 2 Chapter 17 Specification Contents ...
Page 342: ...17 14 Chapter 17 Specification Technical Information ...
Page 347: ......
Page 348: ......
Page 349: ......