6-16
Chapter 6 Admit/Discharge
Suspend Monitoring
To Resume Monitoring Automatically
The auto resume monitoring function can be used when the patient temporarily leaves the bed.
When the specified condition for ECG, SpO
2
, etc. is met, the monitoring will automatically resume after the
preprogrammed duration.
The following initial settings are required in advance.
Auto Resume ON/OFF
Auto Resume Disable Duration (OFF/5 min./10 min./15 min.)
Resume even during Lead-Off (Enable/Disable)
Auto Resume Duration (10 sec./30 sec./1 min./2 min./3 min./4 min./5 min.)
REFERENCE
Refer to
Maintenance Manual "Other" P7-10 for procedure on initial settings.
Condition for Auto Resume Monitoring
The operation from suspend monitoring to automatically resume monitoring is explained below.
The monitoring will not automatically resume until the set duration for "Auto Resume Disable Duration" (OFF/
5 min./10 min./15 min.) is elapsed.
The monitoring will resume after the set "Auto Resume Disable Duration" elapses, and when the following
condition is met for the set "Auto Resume Duration" (10 sec./30 sec./1 min./2 min./3 min./4 min./5 min.).
NOTE
When the [Resume] key is pressed during the auto resume duration, the [Resume] key
operation will be prioritized.
Condition to Automatically Resume Monitoring
The monitoring will resume when the telemetry reception condition is good, and either one of ECG, RESP, SpO
2
, SpO
2
-2, NIBP,
BP1, BP2, TEMP1, TEMP2, CO
2
meets the specified condition.
Measurement
Condition to Resume Monitoring
Telemetry Reception
The reception condition is good.
ECG
[Enable] is set for "Resume even during Lead-Off": Resumes monitoring even during the lead-
off condition when the measurement is started.
[Disable] is set for "Resume even during Lead-Off": Monitoring will not resume during the
lead-off condition even if the measurement is started.
RESP
[Enable] is set for "Resume even during Lead-Off": Resumes monitoring even during the lead-
off condition when the measurement is started.
[Disable] is set for "Resume even during Lead-Off": Monitoring will not resume during the
lead-off condition even if the measurement is started.
SpO
2
, SpO
2
-2
The measurement is started, and SpO
2
disconnected condition is not detected.
NIBP
The measurement data is updated.
BP
The measurement is started.
TEMP1, TEMP2
The measurement is started.
CO
2
The measurement is started.
Summary of Contents for DYASCOPE 8000 Series
Page 1: ......
Page 2: ......
Page 12: ...Contents ...
Page 20: ...Contents ...
Page 44: ...1 2 Chapter 1 General Description Contents ...
Page 50: ...2 2 Chapter 2 Name of Parts and Their Functions Contents ...
Page 54: ...2 4 Chapter 2 Name of Parts and Their Functions ...
Page 56: ...3 2 Chapter 3 Description of the Display Contents ...
Page 70: ...4 2 Chapter 4 Basic Operation Contents ...
Page 82: ...5 2 Chapter 5 Preparation Contents ...
Page 88: ...5 6 Chapter 5 Preparation Nurse Call Daily Check ...
Page 90: ...6 2 Chapter 6 Admit Discharge Contents ...
Page 110: ...7 2 Chapter 7 Alarm Function Contents ...
Page 138: ...8 2 Chapter 8 Parameter Setup Contents ...
Page 164: ...8 26 Chapter 8 Parameter Setup Parameter ON OFF ...
Page 166: ...9 2 Chapter 9 Data Review Contents ...
Page 190: ...10 2 Chapter 10 Waveform Review Contents ...
Page 206: ...11 2 Chapter 11 Calculation Contents ...
Page 212: ...12 2 Chapter 12 Printing Contents ...
Page 230: ...13 2 Chapter 13 Menu Items Contents ...
Page 270: ...13 40 Chapter 13 Menu Items Nurse Team Setup ...
Page 272: ...14 2 Chapter 14 Troubleshooting Contents ...
Page 296: ...14 24 Chapter 14 Troubleshooting Troubleshooting ...
Page 298: ...15 2 Chapter 15 Setup Item Default Value Contents ...
Page 320: ...15 22 Chapter 15 Setup Item Default Value Menu Central Monitor Display ...
Page 322: ...16 2 Chapter 16 Accessories Contents ...
Page 326: ...16 4 Chapter 16 Accessories Optional Accessories ...
Page 328: ...17 2 Chapter 17 Specification Contents ...
Page 342: ...17 14 Chapter 17 Specification Technical Information ...
Page 347: ......
Page 348: ......
Page 349: ......