Chapter 14 Troubleshooting
Troubleshooting
General
The bedside monitor is not properly
connected to the DS-LANIII network.
Connect using the Ethernet branch cable (CJ-522) or
LAN interface cable (CJ-530).
The power is turned OFF on the bedside
monitor or telemetry receiver.
Turn ON the power.
The software version of the telemetry
receiver does not correspond.
Contact your nearest service representative.
The ECG waveform and HR are not
displayed.
[BP] is selected for HR alarm source on
the DS-7000 series bedside monitor.
Select [HR]/ [SpO
2
] for HR alarm source.
The BP numeric data, BP
waveform, NIBP numeric data are
not displayed.
The BP measurement unit (mmHg/kPa)
is different between the bedside monitor
and the central monitor.
If the measurement unit is different between the
bedside monitor and the central monitor, BP
waveform, BP numeric data, NIBP numeric data will
not be transmitted from the bedside monitor.
In such case, it will be treated as not measured data,
and will not be displayed on the central monitor. Make
sure to set the same measurement unit for the bedside
monitor and the central monitor.
(
Maintenance Manual "Unit" P7-9)
The temperature numeric data is
not displayed.
The temperature measurement unit (°C/
°F) is different between the bedside
monitor and the central monitor.
If the temperature measurement unit is different
between the bedside monitor and the central monitor,
the temperature data will not be transmitted from the
bedside monitor.
In such case, it will be treated as not measured data,
and will not be displayed on the central monitor. Make
sure to set the same measurement unit for the bedside
monitor and the central monitor.
(
Maintenance Manual "Unit" P7-9)
Situation
Cause
Solution
The data was initialized when the
power was turned ON.
The power has been turned OFF for
more than 5 minutes.
When the power has been turned OFF for more than 5
minutes, NIBP list will be deleted. To retain the data,
turn ON the power within 5 minutes.
Internal memory error is generated.
Contact your nearest service representative.
The data is initialized each time the
power is turned ON.
The internal switch is set to initialize.
The setting of the rotary switch needs to be changed.
Contact your nearest service representative.
The backup battery is depleted.
The long-term backup battery needs to be
replaced.Contact your nearest service representative.
The display is too dark.
The display brightness is not
adjusted.
Due to the LCD characteristic, the visible range is
limited.
Adjust to the appropriate brightness on the "Brightness
Setup" screen.
(
The "Check Backup Battery"
message is displayed.
The battery for the backup memory is
depleted.
The long-term backup battery needs to be
replaced.Contact your nearest service representative.
There is an offset in the touch
panel.
The detecting location is misaligned
due to change over time.
Calibration needs to be performed. Contact your
nearest service representative.
The touch panel does not function
properly.
A scratch on the touch panel surface
or foreign object entering the touch
panel junction is causing misdetection
of the key area.
The touch panel needs to be replaced. Contact your
nearest service representative.
The date/time is displayed in
yellow.
The time synchronization with the
SNTP server or patient data server
has failed.
Check the connection with the SNTP server or patient
data server.
Situation
Cause
Solution
Summary of Contents for DYASCOPE 8000 Series
Page 1: ......
Page 2: ......
Page 12: ...Contents ...
Page 20: ...Contents ...
Page 44: ...1 2 Chapter 1 General Description Contents ...
Page 50: ...2 2 Chapter 2 Name of Parts and Their Functions Contents ...
Page 54: ...2 4 Chapter 2 Name of Parts and Their Functions ...
Page 56: ...3 2 Chapter 3 Description of the Display Contents ...
Page 70: ...4 2 Chapter 4 Basic Operation Contents ...
Page 82: ...5 2 Chapter 5 Preparation Contents ...
Page 88: ...5 6 Chapter 5 Preparation Nurse Call Daily Check ...
Page 90: ...6 2 Chapter 6 Admit Discharge Contents ...
Page 110: ...7 2 Chapter 7 Alarm Function Contents ...
Page 138: ...8 2 Chapter 8 Parameter Setup Contents ...
Page 164: ...8 26 Chapter 8 Parameter Setup Parameter ON OFF ...
Page 166: ...9 2 Chapter 9 Data Review Contents ...
Page 190: ...10 2 Chapter 10 Waveform Review Contents ...
Page 206: ...11 2 Chapter 11 Calculation Contents ...
Page 212: ...12 2 Chapter 12 Printing Contents ...
Page 230: ...13 2 Chapter 13 Menu Items Contents ...
Page 270: ...13 40 Chapter 13 Menu Items Nurse Team Setup ...
Page 272: ...14 2 Chapter 14 Troubleshooting Contents ...
Page 296: ...14 24 Chapter 14 Troubleshooting Troubleshooting ...
Page 298: ...15 2 Chapter 15 Setup Item Default Value Contents ...
Page 320: ...15 22 Chapter 15 Setup Item Default Value Menu Central Monitor Display ...
Page 322: ...16 2 Chapter 16 Accessories Contents ...
Page 326: ...16 4 Chapter 16 Accessories Optional Accessories ...
Page 328: ...17 2 Chapter 17 Specification Contents ...
Page 342: ...17 14 Chapter 17 Specification Technical Information ...
Page 347: ......
Page 348: ......
Page 349: ......