FUSION SPLICER 90S Rev5
102
(2)
Cannot change the parameters of the RS02/03 or CT50 with the fusion splicer.
・
The CT50 may be already connected to another fusion splicer. Push and hold the
Link button on the device. This will cause the Link LED of the device to blink and
start the paring process. After this, try to connect with the stripper. Refer to
Wireless data communication section.
(3)
Cannot connect to the smartphone.
・
Check the specification of the smartphone.The required specifications are as
follows.
The Bluetooth wireless communication function of the splicer is not guaranteed
to work with all smartphones.
Wireless communication technology: Bluetooth 4.1 or later
OS version for iPhone: iOS8.0 or later (after iPhone6)
Android: Android
5.0 or later
Supervising
(1)
What functions can be disabled
・
See Section [Menu Lock Settings].
(2)
Method to lock “selection” or “editing” of Splice or Heater mode
・
See Section [Menu Lock Settings ].
(3)
Method to set parameters of Splice or Heater mode from a PC
・
Refer to the communication software “Data Connection” included on fusion
splicer hard-drive.
(4)
Forgot password
・
Contact the nearest Fujikura authorized distributor.
Other Functions
(1)
Method to hide messages on [READY] screen
・
Change the fiber image from X/Y view to X magnified view or Y magnified view
by pressing X/Y key.
(2)
Too many repetitions until “Test Finish” indicated in [Arc Calibration]
・
The fusion splicer needs to repeat the arc calibration after replacing the
electrodes or when the environmental conditions change drastically. The
number of arc calibrations can be set to a specific amount. When the fusion
splicer completes the set amount of calibrations, it indicates “Test Finish”.
However, this does not mean it is completely calibrated.
(3)
“Test Finish” is never indicated after many repetitions in [Arc Calibration]
・
Execute [Stabilize Electrodes] function in [Maintenance Menu]. If the fusion
splicer still does not indicate “Test Finish”, replace the electrodes. See section
[Replace Electrodes] function.
(4)
No arc power change after [Arc Calibration]
・
An internal factor is calibrated and adjusted for the specific arc power selected.
The displayed arc power in each splice mode does not change.
・
The calibration results affect all splice modes.
(5)
Method to input different comments after each splice in splice results data
・
See section [Splice Memory Comment].
(6)
Method to download splice results from fusion splicer to PC
・
Refer to the instruction manual “Data Connection” included in the splicer.
Press [F1] key after installing the software to display the instruction manual.
Contact the the nearest Fujikura authorized distributor for further inquiries.
Trademarks
The following logos and words appearing in this manual are trademarks or
registered trademarks of their respective owners. Any use of such marks by Fujikura
Ltd is under license.
・
Bluetooth
®
word mark and logos are the registered trademarks of Bluetooth SIG,
Inc.
・
iPhone
®
is the registered trademarks of Apple Inc.
・
iOS
®
is a registered trademark of Cisco Systems Inc. of the United States.
・
Android
®
is the registered trademarks of Google Inc.