plice loss increase: Cause and remedy (1 of 2)
Symptom
Cause
Remedy
Axis offset
Dust on v-groove
or v-groove clamp
chip
Clean V-groove and v-groove
clamp chip.
Fiber angle
Dust on v-groove
or v-groove clamp
chip
Clean v-groove and v-groove
clamp chip.
Bad fiber end-
face quality
Check if fiber cleaver is well
conditioned.
Core step
Dust on v-groove
or v-groove clamp
chip
Clean v-groove and v-groove
clamp chip.
Core curve
Bad fiber end-
face quality.
Check if fiber cleaver is well
conditioned.
Prefuse power too
low or prefuse
time too short.
Increase [Prefuse Power] and/or
[Prefuse Time].
MFD mismatch
Arc power too
low
Increase [Arc Power] and/or
[Arc Time].
Combustion
Bad fiber end-face
quality
Check the cleaver
Dust still present
after cleaning fiber
or cleaning arc.
Clean fiber thoroughly or Increase
[Cleaning Arc Time]
Bubbles
Bad fiber end-face
quality
Check if fiber cleaver is well
conditioned.
Prefuse power too
low or prefuse
time too short.
Increase [Prefuse Power] and/or
[Prefuse Time].
Splice loss increase: Cause and remedy (2 of 2)
Symptom
Cause
Remedy
Separation
Fiber stuffing too
small
Perform [Motor Calibration]
Prefuse power too
high or prefuse
time too long.
Decrease [Prefuse Power] and/or
[Prefuse Time].
Fat
Fiber stuffing too
much
Decrease [Overlap] and perform
[Motor Calibration].
Thin
Arc power not
adequate
Perform [Arc Calibration].
Some arc
parameters not
adequate
Adjust [Prefuse Power], [Prefuse
Time] or [Overlap].
Line
Some arc
parameters not
adequate
Adjust [Prefuse Power], [Prefuse
Time] or [Overlap].
A vertical line sometimes appears at the splice point when MM fibers
or dissimilar fibers (different diameters) are spliced. This does not
affect splice quality, such as splice loss or tensile strength.
A loss becomes high when the following errors occur.
Electric discharge may become unstable when usage environment
changes significantly. In such a case, please use it after performing
“Stabilize Electrodes”.
When SET key has been
pushed, "Stabilize
Electrodes" is performed.
Please perform an "Arc
Calibration" after the end
of "Stabilize Electrodes"
according to the directions
on a screen.