FUSION SPLICER 90S Rev5
23
DC operation with external battery
1.
When operating the fusion splicer with DC power, use the Power Unit,
Battery Pack (BTR-15) or DC supply (DCA-03), and DC cord (DCC-20 or
DCC-21).
2.
Attach the power unit into the fusion splicer body and put the cover on it.
3.
Connect the DC cord to the DC supply.
4.
Insert the DC cord into the inlet of the Power Unit.
Battery Operation
Please use the supplied battery pack BTR-15.
Before using the battery
Check the remaining battery capacity. If it is 20% or less before operation, you will
only be able to use the machine for long.
Do not supply DC15V or greater.
If the polarity (positive / negative) is wrong, the protection circuit in
DCA-03 or BTR-15 shuts down the DC output.
DC cord
[DCC-20]
DC cord
[DCC-21]
Car Cigar socket to BTR-15 or DCA-03
Car battery to BTR-15 or DCA-03
Adhere to below practices to prevent battery damage.
The charge of the battery pack after air transport may be 30% or
less in accordance with the aviation law. When it arrives, please
charge the battery pack to full.
The capacity of the battery gradually decreases even if it is not
used. If the battery discharges completely, the battery may no
longer be able to be re-charged. Charge the battery before
storing for a long time and after every use.
If a battery is to be stored for a long time, periodical charge
every six months is recommended regardless of battery charge
level.
Follow below operating conditions
Operation
: -10 degree C ~ 50 degree C
Charging
: 0 degree C ~
4
0 degree C
Storage
: -20 degree C ~ 30 degree C