Steer the distal end of the endoscope to the region of interest by turning the up/down and left/
right angulation knobs.
Note
To retain the bending angle securely, hold the up/down and left/right angulation knobs
by hand. When the up/down and left/right angulation knobs are not held, even if the
up/down and left/right angulation knobs are locked with the up/down and left/right
angulation locks, the bending angle at the distal end may change due to advancement
or withdrawal of the endoscope or insertion of an endotherapy device into the
instrument channel.
Locking of Up/Down Angulation Lock
Locking of Left/Right Angulation Lock
(2)
When necessary, retain the bending angle of the distal end by moving the up/down and left/
right angulation locks in the opposite direction of F (Free).
Note
When moving the up/down and left/right angulation locks, hold the up/down and left/
right angulation knobs by hand. When the up/down and left/right angulation knobs are
not held, the bending angle at the distal end may change.
ED580XT_E2-30_897N200877B.indb 8
2018/09/20 15:50:24
Summary of Contents for ED-580XT
Page 2: ...ii 897N200877B ED580XT_E2 30_897N200877B indb 2 2018 09 20 15 50 15...
Page 12: ...897N200877B 4 Introduction ED580XT_E2 30_897N200877B indb 4 2018 09 20 15 50 16...
Page 32: ...897N200877B 1 20 Chapter 1 Precautions ED580XT_E2 30_897N200877B indb 20 2018 09 20 15 50 17...
Page 52: ...897N200877B 3 4 Chapter 3 Workflow ED580XT_E2 30_897N200877B indb 4 2018 09 20 15 50 19...
Page 116: ...897N200877B 7 2 Chapter 7 Service ED580XT_E2 30_897N200877B indb 2 2018 09 20 15 50 25...
Page 134: ...897N200877B Appendix 18 Appendix ED580XT_E2 30_897N200877B indb 18 2018 09 20 15 50 27...