897N200877B
4-20
Chapter 4 Preparation and Inspection
4.5 Preparing Related Equipment
Prepare related equipment such as the following as per instructions provided in each operation
manual.
• Light source
• Processor
• Monitor
• Water tank
• Suction unit
• Endotherapy device
• Mouthpiece
• Endoscopic CO
2
Regulator
• Electrosurgical unit
4.5.1 Inspecting Related Equipment
Inspect related equipment as per instructions provided in each operation manual.
Note
Some pieces of related equipment are not reprocessed before shipping from FUJIFILM
or the manufacturer. Reprocess them for the first time prior to use as per instructions
provided in respective operation manuals.
4.5.2 Preparing System
CAUTION
• With regard to the amount of sterile water in the water tank, follow the instructions
provided in the operation manual of the water tank. If the amount of sterile water in the
water tank exceeds the limit, the air/water supply function may be disabled or it may cause
equipment failure due to contact with spilled water.
• Use the endoscopic CO
2
regulator described in this manual. If another insufflation device
is connected, the air/water supply function lessens and may result in improper cleaning of
the lens.
(1)
Move the cart with the processor, light source and other related equipment to the place where
endoscope is to be used.
Note
• Refer to the Operation Manual of the processor/light source to install related
equipment onto the cart.
• When multiple foot switches are used, check the position of the corresponding foot
switch in advance, so as not to use the wrong foot switch by mistake.
(2)
After turning the main switch on the cart to OFF position, insert the AC plug of the cart into a
hospital grade receptacle.
ED580XT_E2-30_897N200877B.indb 20
2018/09/20 15:50:22
Summary of Contents for ED-580XT
Page 2: ...ii 897N200877B ED580XT_E2 30_897N200877B indb 2 2018 09 20 15 50 15...
Page 12: ...897N200877B 4 Introduction ED580XT_E2 30_897N200877B indb 4 2018 09 20 15 50 16...
Page 32: ...897N200877B 1 20 Chapter 1 Precautions ED580XT_E2 30_897N200877B indb 20 2018 09 20 15 50 17...
Page 52: ...897N200877B 3 4 Chapter 3 Workflow ED580XT_E2 30_897N200877B indb 4 2018 09 20 15 50 19...
Page 116: ...897N200877B 7 2 Chapter 7 Service ED580XT_E2 30_897N200877B indb 2 2018 09 20 15 50 25...
Page 134: ...897N200877B Appendix 18 Appendix ED580XT_E2 30_897N200877B indb 18 2018 09 20 15 50 27...