Chapter 4
Prepara
tion and Inspectio
n
<Inspection>
Visually check that the suction valve is free from abnormalities such as tears, distortions,
cracks, indentations, etc.
Note
The suction valve is a consumable item. If any abnormality is found, use a new
reprocessed valve.
<Attachment>
Note
The suction valve and the suction valve cylinder have an orange marking. Check the
marking on the suction valve and that on the suction valve cylinder to prevent an error
in attachment.
Bulge
Rotation
Metal
Ridges
Holes
Orange
marking
Suction Valve Cylinder
Top View
Suction Valve
Bottom View
Orange marking
(1)
Attach the suction valve to the endoscope’s suction valve cylinder, align the two metal ridges of
the suction valve with the two holes in the suction valve cylinder, and push in the valve firmly.
(2)
Visually and manually check that the valve fits properly without any bulging of the skirt. Also
confirm that the valve cannot be rotated.
Forceps Valve
Suction Valve
Air/Water Valve Distal End Cap
4.4 Inspecting and Attaching Accessories
ED580XT_E2-30_897N200877B.indb 15
2018/09/20 15:50:22
Summary of Contents for ED-580XT
Page 2: ...ii 897N200877B ED580XT_E2 30_897N200877B indb 2 2018 09 20 15 50 15...
Page 12: ...897N200877B 4 Introduction ED580XT_E2 30_897N200877B indb 4 2018 09 20 15 50 16...
Page 32: ...897N200877B 1 20 Chapter 1 Precautions ED580XT_E2 30_897N200877B indb 20 2018 09 20 15 50 17...
Page 52: ...897N200877B 3 4 Chapter 3 Workflow ED580XT_E2 30_897N200877B indb 4 2018 09 20 15 50 19...
Page 116: ...897N200877B 7 2 Chapter 7 Service ED580XT_E2 30_897N200877B indb 2 2018 09 20 15 50 25...
Page 134: ...897N200877B Appendix 18 Appendix ED580XT_E2 30_897N200877B indb 18 2018 09 20 15 50 27...