Chapter 1 Precautions
WARNING
• Do not forcibly advance or withdraw the endoscope into/from the patient, angulate the
bending section forcibly or operate it quickly. It may cause patient injury, bleeding or
perforation.
• Insert the endoscope while observing the endoscopic image to secure patient safety. Not
doing so may cause patient pain, injury, bleeding and/or perforation.
• Do not forcibly turn the angulation knob further after turning the knob until it stops. If the
angulation knob is forcibly turned, the angulation mechanism may malfunction and the
bending section may not return to its neutral position, making it difficult to withdraw the
endoscope.
• If the bending section does not return to its neutral position during a procedure, do not
withdraw the endoscope forcibly. Consult your local FUJIFILM dealer. If the endoscope
is withdrawn forcibly, injury to tissues in the body cavity, bleeding and/or perforation may
result.
• Be extremely careful when performing retroflexed observation in a narrow lumen. Do
not perform retroflexed observation forcibly. Otherwise, it may become impossible to
straighten the angle of the bending section and/or withdraw the endoscope from the
patient.
• Avoid aspirating solid materials or thick fluids. If the suction valve does not return to its
original position, stop aspiration immediately and slowly withdraw the endoscope. If any
solid materials or thick fluids adhere to or clog the suction valve, suction may not stop,
causing damage to mucous membrane.
• When injecting fluids by attaching a syringe to the forceps valve, open the lid of the
forceps valve and insert the syringe straight into the forceps valve. Otherwise, the forceps
valve may be damaged or the syringe may be accidentally detached during fluid injection
and body fluids may leak or splash from the forceps valve, posing an infection control risk
to the patient or end-user.
• Do not strongly press the endotherapy device against tissues in the body cavity.
Otherwise, injury to tissues in the body cavity, bleeding and/or perforation may result.
• Unless the endoscope's bending section is in a neutral position (essentially "straight"), do
not withdraw the endoscope whenever the endoscopic image is compromised, unclear,
blurry, etc. due to any reason or condition including loss of image, power interruption,
water droplets or dirt/debris adhering to the objective lens, etc. Doing so may cause injury
to tissues in the body cavity, bleeding and/or perforation.
• Firmly connect the LG connector of the endoscope and the light source as well as the
video connector and the processor. If the LG connector is not connected properly, the
endoscopic image may flicker or be lost, which may cause patient injury, bleeding and/or
perforation.
• Firmly connect the video connector of the endoscope and the processor. If the video
connector is not connected properly, the endoscopic image may flicker or be lost, which
may cause patient injury, bleeding and/or perforation.
ED580XT_E2-30_897N200877B.indb 12
2018/09/20 15:50:17
Summary of Contents for ED-580XT
Page 2: ...ii 897N200877B ED580XT_E2 30_897N200877B indb 2 2018 09 20 15 50 15...
Page 12: ...897N200877B 4 Introduction ED580XT_E2 30_897N200877B indb 4 2018 09 20 15 50 16...
Page 32: ...897N200877B 1 20 Chapter 1 Precautions ED580XT_E2 30_897N200877B indb 20 2018 09 20 15 50 17...
Page 52: ...897N200877B 3 4 Chapter 3 Workflow ED580XT_E2 30_897N200877B indb 4 2018 09 20 15 50 19...
Page 116: ...897N200877B 7 2 Chapter 7 Service ED580XT_E2 30_897N200877B indb 2 2018 09 20 15 50 25...
Page 134: ...897N200877B Appendix 18 Appendix ED580XT_E2 30_897N200877B indb 18 2018 09 20 15 50 27...